Drift Test

  • Align microscope
  • 200kx recommended magnification
  • Insert quantifoil grid
  • Insert holder
  • Cool down holder if testing cryo holder
  • In Tvips software
    • Open EMMENU
    • In the Window menu open Output log (F2)
    • select "Drift" button on EMMENU panel
    • 30 images in 30 second increments
    • at the end of the test select "store" on the output window
    • Then "clear" output window
  • Repeat test 3 times