F20 Daily Checklist

  • Cool microscope cold finger
  • Insert holder to first stop
    • Do not insert holder
  • Open colum valves
    • check for beam

  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
  • 50 micron condenser aperture (check white board, for now it's position #2)
Deleted:
<
<
  • spot size 2
 

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
Changed:
<
<
  • Adjust the gun tilt alignment to Trash.findDFdf the brightest spot
>
>
  • Adjust the gun tilt alignment to locate the brightest spot
 
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

-- KdDerr - 10 Oct 2006

Revision 1205 Jul 2011 - Main.MarienaSilvestry

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • Insert holder to first stop
    • Do not insert holder
  • Open colum valves
    • check for beam

  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
  • 50 micron condenser aperture (check white board, for now it's position #2)
  • spot size 2

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
  • Adjust the gun tilt alignment to Trash.findDFdf the brightest spot
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

-- KdDerr - 10 Oct 2006

Revision 1105 Jul 2011 - Main.MarienaSilvestry

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • Insert holder to first stop
    • Do not insert holder
  • Open colum valves
    • check for beam
Deleted:
<
<
 
  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
Changed:
<
<
  • 50 micron condenser aperture
>
>
  • 50 micron condenser aperture (check white board, for now it's position #2)
 
  • spot size 2

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
  • Adjust the gun tilt alignment to Trash.findDFdf the brightest spot
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

-- KdDerr - 10 Oct 2006

Revision 1017 Sep 2008 - Main.DavidCowburn

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • Insert holder to first stop
    • Do not insert holder
  • Open colum valves
    • check for beam

  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
  • 50 micron condenser aperture
  • spot size 2

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
  • Adjust the gun tilt alignment to Trash.findDFdf the brightest spot
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

-- KdDerr - 10 Oct 2006

Revision 916 Jul 2008 - Main.DavidCowburn

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • Insert holder to first stop
    • Do not insert holder
  • Open colum valves
    • check for beam

  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
  • 50 micron condenser aperture
  • spot size 2

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
Changed:
<
<
  • Adjust the gun tilt alignment to find the brightest spot
>
>
  • Adjust the gun tilt alignment to Trash.findDFdf the brightest spot
 
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

-- KdDerr - 10 Oct 2006

Revision 823 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
Changed:
<
<
  • insert holder to first stop
>
>
  • Insert holder to first stop
 
    • Do not insert holder
Deleted:
<
<
  • Uncheck Turbo Auto Off
 
  • Open colum valves
    • check for beam
Deleted:
<
<
  • Operation Parameters * Arround 3900-3950 extraction voltage * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2
 
  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using
Changed:
<
<
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
      • 2 sec exposure means 1 nanoAmp of current
>
>

Operation Parameters for alignments

  • 3850-3950 extraction voltage
  • Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.)
  • 50 micron condenser aperture
  • spot size 2
 
Changed:
<
<
  • Condensor Stigmation
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle
>
>

Gun tilt (check)

  • Magnification 10kx
  • Check exposure time without the small screen
    • Around 2 sec is appropriate
    • 2 sec exposure means 1 nanoAmp of current
      • if the exposure time is not close to 2 seconds double check for correct operation parameters
      • If the operation parameters are correct adjust Gun Tilt
  • Location: tune/alignments/gun
  • Adjust the gun tilt alignment to find the brightest spot
  • After adjusting Gun tilt:
    • If exposure time is over 2 sec increase the extraction voltage and readjust gun tilt
Added:
>
>
    • If exposure time is less than 1.4 sec lower the extraction voltage and readjust gun tilt
 
Deleted:
<
<
  • Flatfield Tietz camera
 
Changed:
<
<
  • New user Low Dose setup
    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file listed on whiteboard
    • save as new user file and folder
>
>

Condensor Stigmation

  • Spot size 2
  • Magnification 250kx
  • Condense beam to 30mm size
  • Press stigmator button
Added:
>
>
    • select condensor
  • Use multifunction knobs to make beam round

Check image of FEG Tip

  • Condense beam to point
    • Beam may not be round at crossover, this is the image of the FEG tip
    • 5mm circle = 10nm
    • You want the majority of your beam to fit within the 5mm circle

Flatfield Tietz camera

 

-- KdDerr - 10 Oct 2006

Revision 723 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam

  • Operation Parameters * Arround 3900-3950 extraction voltage * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2

  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using

  • Checking Gun Tilit/Brightness
    • Magnification 10kx
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
      • 2 sec exposure means 1 nanoAmp of current

  • Condensor Stigmation
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle

  • Flatfield Tietz camera

  • New user Low Dose setup
    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
Changed:
<
<
    • Load known good Low Dose listed on whiteboard
>
>
    • Load known good Low Dose file listed on whiteboard
 
    • save as new user file and folder

-- KdDerr - 10 Oct 2006

Revision 617 Jul 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam

  • Operation Parameters
Changed:
<
<
* 3950 extraction voltage
>
>
* Arround 3900-3950 extraction voltage
  * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2
Added:
>
>
  • Start Low Dose and click through systematically to "exposure" mode
    • Load Lose Dose file you will be using
 
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
      • 2 sec exposure means 1 nanoAmp of current

  • Condensor Stigmation
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle
Added:
>
>
  • Flatfield Tietz camera
 
  • New user Low Dose setup
    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
Changed:
<
<
    • Load known good Low Dose file like KD0.lds
>
>
    • Load known good Low Dose listed on whiteboard
 
    • save as new user file and folder

-- KdDerr - 10 Oct 2006

Revision 501 Jun 2007 - Main.DavidCowburn

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam

  • Operation Parameters * 3950 extraction voltage * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
      • 2 sec exposure means 1 nanoAmp of current

  • Condensor Stigmation
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle

  • New user Low Dose setup
    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file like KD0.lds
    • save as new user file and folder
Deleted:
<
<

  • Set ALLOWTOPICVIEW =
 -- KdDerr - 10 Oct 2006

Revision 410 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

  • Cool microscope cold finger
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam

  • Operation Parameters * 3950 extraction voltage * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
      • 2 sec exposure means 1 nanoAmp of current

  • Condensor Stigmation
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle
Added:
>
>
  • New user Low Dose setup
    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file like KD0.lds
    • save as new user file and folder
 

  • Set ALLOWTOPICVIEW =

-- KdDerr - 10 Oct 2006

Revision 310 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

F20 Daily Checklist

Added:
>
>
 
  • Cool microscope cold finger
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam
Added:
>
>
 
  • Operation Parameters
Changed:
<
<
o 3950 extraction voltage o Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) o 50 micron condenser aperture o spot size 2
>
>
* 3950 extraction voltage * Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) * 50 micron condenser aperture * spot size 2
 
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
Added:
>
>
    • Check exposure time without the small screen
      • Around 2 sec is appropiate
 
      • 2 sec exposure means 1 nanoAmp of current

  • Condensor Stigmation
Changed:
<
<
o spot size 2 o Magnification 250kx o condense beam to 30mm size o press stigmator button o select condensor stigmation o use multifunction knobs to make beam round o Condense beam to point + beam may not be round at crossover, this is the image of the FEG tip + 5mm circle = 10nm + you want hte majority of your beam to fit within the 5mm circle
>
>
    • spot size 2
    • Magnification 250kx
    • condense beam to 30mm size
    • press stigmator button
    • select condensor stigmation
    • use multifunction knobs to make beam round
    • Condense beam to point
      • beam may not be round at crossover, this is the image of the FEG tip
      • 5mm circle = 10nm
      • you want the majority of your beam to fit within the 5mm circle
Added:
>
>
 
  • Set ALLOWTOPICVIEW =

-- KdDerr - 10 Oct 2006

Revision 210 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents
Changed:
<
<
  • Operation Parameters
>
>

F20 Daily Checklist

  • Cool microscope cold finger
Added:
>
>
  • insert holder to first stop
    • Do not insert holder
  • Uncheck Turbo Auto Off
  • Open colum valves
    • check for beam
  • Operation Parameters
  o 3950 extraction voltage o Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) o 50 micron condenser aperture o spot size 2
Changed:
<
<
* Checking Gun Tilit/Brightness o Magnification 10kx + 2 sec exposure means 1 nanoAmp of current
>
>
  • Checking Gun Tilit/Brightness
    • Magnification 10kx
      • 2 sec exposure means 1 nanoAmp of current
 
Changed:
<
<
* Condensor Stigmation o spot size 2 o Magnification 250kx o condense beam to 30mm size o press stigmator button o select condensor stigmation o use multifunction knobs to make beam round o Condense beam to point + beam may not be round at crossover, this is the image of the FEG tip + 5mm circle = 10nm + you want hte majority of your beam to fit within the 5mm circle
>
>
  • Condensor Stigmation o spot size 2 o Magnification 250kx o condense beam to 30mm size o press stigmator button o select condensor stigmation o use multifunction knobs to make beam round o Condense beam to point + beam may not be round at crossover, this is the image of the FEG tip + 5mm circle = 10nm + you want hte majority of your beam to fit within the 5mm circle
 
  • Set ALLOWTOPICVIEW =

-- KdDerr - 10 Oct 2006

Revision 110 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

  • Operation Parameters o 3950 extraction voltage o Gun Lens 3 (make sure it is gun lens 3 by toggling between 2 and 3.) o 50 micron condenser aperture o spot size 2 * Checking Gun Tilit/Brightness o Magnification 10kx + 2 sec exposure means 1 nanoAmp of current

* Condensor Stigmation o spot size 2 o Magnification 250kx o condense beam to 30mm size o press stigmator button o select condensor stigmation o use multifunction knobs to make beam round o Condense beam to point + beam may not be round at crossover, this is the image of the FEG tip + 5mm circle = 10nm + you want hte majority of your beam to fit within the 5mm circle