Instructions for finding the beam
Still cannot Trash.findDFdf beam
2100 only
3200 only
1230 only
Instructions for finding the beam
Move sample to make sure you are not on a grid bar
If you are using the cryostage make sure the shutter is open
Make sure High Tension is on
Make sure the colum valves are open
Still cannot Trash.findDFdf beam
Slowly drop the magnification one step at a time
Once beam is located
Condense beam
Center the spot
2100 only
Press "photo" button
microscope may think it is taking a picture
3200 only
Make sure energy filter is not inserted
1230 only
Make sure camera is not inserted