Insert sample
Line up pin on holder
Insert sample to first sample stop (marked line indicates correct depth)
ID holder and hit return (bottom left)
Watch for red light to go out
Turn holder slowly anti-clockwise to stop
Slowly allow the vacuum to pull the holder in
--
DavidStokes
- 18 Mar 2005
Set ALLOWTOPICCHANGE =
CemfacGroup
>
>
Return to Cryo EM Website at
http://www.nysbc.org/facilities/CEM
Insert sample
Open Vacuum Overview page
Prepump Airlock
Line up pin on holder
Insert sample to first sample stop (marked line indicates correct depth)
ID holder in TUI and hit return (bottom left)
Watch specimen pumpdown timer go to zero on Vacuum overview page
Turn holder slowly anti-clockwise to stop
Apply resistance and slowly allow the vacuum to pull the holder in
--
DavidStokes
- 18 Mar 2005
Set ALLOWTOPICCHANGE =
CemfacGroup