Microscope Parameters
wavelengths
objective lenses
Aperture Sizes
CCD information
Microscope Parameters
wavelengths
80kV: 0.04176 A
120kV: 0.03349 A
objective lenses
f = ??
Cs = ?? mm
Chromatic aberration = ?? mm
Aperture Sizes
Condensor : 300 um, 200 um, 100 um
Objective : 80 um, 50 um, 20 um
Selected Area: 300 um, 100 um, 20 um
CCD information
Size of camera : 1350 x 1040 pixels
Pixel size: 6.45 um
Pre-film mag : 0.0617 (!!)
Changed:
<
<
Gain uncalibrated
>
>
Gain uncalibrated, unlikely to be used for low-dose ccd acquisition
Magnification Calibration : Jeol 1230
Magnifications calibrated using 2160 cross grid using method described in
SerialEM
documentation
Film Magnification
CCD Magnification
Unbinned Pixel Size (A)
Rotation (relative to mag 1 step lower)
2,500
153
421
NA
3,000
185
349
0.34
4,000
246
262
0.29
5,000
307
210
-0.05
6,000
368
175
-3.03
8,000
491
131
-6.10
10,000
625
103
-6.27
12,000
757
85.2
6.58
15,000
920
70.1
-0.58
20,000
1,230
53.2
0.64
25,000
1550
41.7
0.00
30,000
1830
35.2
0.41
40,000
2470
26.1
0.43
50,000
3080
20.95
0.41
60,000
3710
17.4
0.21
80,000
4960
13.0
0.41
100,000
6210
10.4
-0.12
120,000
7500
8.60
0.45
150,000
9440
6.83
-0.10
200,000
12,570
5.13
0.56
250,000
14,740
4.38
0.47
300,000
18,140
3.56
-2.11
* Overall CCD to film factor: 0.0617
Set ALLOWTOPICVIEW =
--
BillRice
- 28 Sep 2007
View topic
|