User Instructions for 2100F

  • Assumed that the microscope has already been aligned and software set up for imaging
  • All users REQUIRE training with staff before using K2 camera
  • In general, this microscope will be used only for cryo samples
Added:
>
>

General guidelines for leginon use

  • Expect to spend the morning getting leginon set up
  • You should commit to a possible all-night session if grid is good
    • no sample -- basic alignment checks / verification Staff or expert users
    • Carbon sample -- focus checks, astigmatism, coma-free Staff or expert users
    • Load Cryo sample Staff or advanced users
    • Grid image to find square; initial check Staff or experienced users
    • Z-focus to make eucentric Staff or experienced users
    • Grid atlas Staff or experienced users
  • Once grid is good
    • pick squares (1 acquire per square)
      • best pick one square first, then just a few holes on this square. If good, go back to pick more squares
    • Pick good holes
      • maybe best to turn off hole finding routine at first
      • Pick many holes per square, but only one focus position (used for z-height)
      • This step takes awhile
    • Re-pick holes for centering step
      • one acquire point, does not need to be in exact middle
      • no focus point (would be ignored anyway)
      • This step takes ~30 min for a "full" square
    • Finally pick acquire point
      • one point per hole due to beam size
      • add one focus point on carbon for drift monitor / focus

Further Notes on j2100

  • Add a melt step for focus (20 sec exposure)
  • Need to turn on FLC for "gr" preset
    • Load "spot 6 for gr" saved FLC preset
  • Turn off FLC when done gr atlas
    • however, sq is still OK with FLC on. Perhaps leave it on for all Low Mag work?
  • preset alignments seem pretty constant. Only check/change if problems seen
  • Imaging of ~30 holes took 1.5 h (starting from submitting exposure queue)
  • Users strongly encouraged to only use leginon interface unless they are experts
  • Use timer for gun valve on exposure step to be safe

 

Controls needed

  • F1 button -- lift screen
  • Specimen movement trackball
  • Focus knobs
  • Brightness control
  • Def/Stig in PLA adjust, in case beam gets out of line
Changed:
<
<
  • Z height to check eucentricity
>
>
  • Z height to check eucentricity
 
  • Mag 1 button to reset Defocus
Changed:
<
<
  • TEM/CCD toggle to change shuttering control to CCD
>
>
 

iMDS

New Instructions for K2 camera

  • Please read and be familiar with use of Leginon and K2 on EMG website
  • Need to set up a "hysteresis cycle" of Search --> Focus --> MDS Photo --> Search ....
  • Dose (brightness) on all modes should be 10 pA/cm2
  • 10.4 pA/cm2 = 8 e/pixel/second
  • IMPORTANT: Beam MUST be blanked (live mode off) during mode changes, otherwise camera may be hit by bright beam
  • Can run Leginon "Manual" app for movie collection
    • Set Camera Mode to Super resolution
    • Choose to record movies (frame numbers are irrelevant for K2; all are saved)
    • Choose total exposure time and frame exposure time as apprporate
    • Be sure that option to save images is checked (under Advanced Settings)
  • Makes files smaller (raw images saved as bytes, gain correction and alignment done via Appion)

Good settings

  • Search Mode
    • alpha 3 (75C0)
    • 2000X mag
    • To make beam dim enough, change 'Custom" to on and spot size FFFF
  • Focus
    • alpha 2 (95D0)
    • 80,000X mag
    • Spot 5
    • Beam condensed to just cover camera (only viewing center 1/2 or 1/4)
  • Photoset
    • alpha 2 (95D0)
    • 25,000X mag (~1.4 A/pixel)
    • Spot 3-5 (beam very condensed to just cover camera)

  • Use DM for live search and exposure
    • Search: Full CCD, 4X binning
    • Focus: Center 1/2 or 1/4, 1X binning
  • Use Leginon for exposure
    • images to database
    • 3/13/15 -- still saved on local disk, need to be manually transferred to bnc14

Sequence of events

  • Need to set up a "hysteresis cycle" of Search --> Focus --> MDS Photo --> Search ....
  • In Search mode, lift screen and view display with live view in DM (K2 camera) -- Use Search setting
  • Find and center area of interest (use PLA to align Search and Photoset)
  • Stop live viewing
  • Go to Focus mode, change DM preset to Focus
  • Put specimen in focus (zero contrast, Thon rings maximized)
  • Press Mag 1 to reset defocus to 0
  • Adjust defocus to desired level
  • Stop live viewing
  • Go to MDS Photo Mode
  • In Leginon software, press camera icon to take and save image
  • Go back to Search mode. Change DM setting to Search

Advice and Common Issues

Eucentricity

  • After you move very far (several grid squares), it is a good idea to check eucentricity, as sample may be bent
  • In Search mode, center an object of interest
  • In JEOL Oprate window, tilt up in 5 degree increments
  • If you are not eucentric, you will see image shift on the live camera window
  • Adjust Z height until object of interest is re-centered (it will never be exactly recentered)
  • Reset tilt angle to 0 using JEOL interface. Should see no (minimal) sample movement

Cannot find Focus beam

  • Focus beam may disappear if you have to adjust defocus a lot (indicating change in specimen height) or if you leave hysteresis cycle
  • Bring it back as follows:
    1. lower screen to see beam
    2. Make sure PLA button is lit
    3. Move beam to center of screen using Def/Stig X and Y knobs (NOT beam shift knobs!)
  • If beam disappears when moving focus position, call a staff member to redo beam shift / image shift alignment

Search and MDS Photo no longer aligned

  • Search and Photo may go out of alignment, especially if you leave the hysteresis cycle
  1. Before realigning anything, cycle through Search --> Focus -- > MDS Photoset a few times
  2. In Search mode, go to a high-contrast and easy to Trash.findDFdf object, like a hunk of ice
  3. Go to Focus, then MDS photoset
  4. Lower screen
  5. Use specimen movement to center the object manually
  6. Go back to Search mode
  7. Be sure that PLA button is lit
  8. Use DEF/Stig X and Y to re-center object

Astigmatism in Focus mode

* Try re-centering Focus using PLA as described above * Do not worry about astigmatism in focus mode. Because of beam shift and inaccuracy of centering beam over camera, you cannot rely on focus to set stigmatism
  • Instead, do this in Photoset mode at 50,000X mag using live Focus mode in DM
  • When happy, write down Objective Stig lens values (in JEOL menu)
  • Stop live mode, adjust brightness and mag to original values
  • go to Search then Focus
  • In Focus mode, adjust Stig lens values to numbers written above (Photoset gets these from Focus, not vice-versa)

Apertures visible in Focus and Search

  • With small condenser aperture needed for K2 , you often see the edge of the objective aperture in Focus and Search modes
  • Can adjust bright tilt to make beam fully visible
  • Note that each mode has its own bright tilt, so this will not affect exposure mode
  • good idea to do a few hsyteresis cycles after changing this

Use STD Photoset to getr low-mag views of grid

  • Go to Standard Photoset mode (normally not used at all in cycle)
  1. Press Low Mag button and lower mag to 150X or lower
  2. manually find and center good squares.
  3. Store positions in JEOL stage interface for later recall
  4. Go back to MDS Photoset when done. Leave for 15 minutes to allow lenses to warm back up

SerialEM

  • 3/15/15 -- not yet installed on K2
  • Never try to use SerialEM and MDS/iMDS at the same time
  • Never use SerialEM with the standard tilt holder

  • Set ALLOWTOPICVIEW =