Maintenance Instructions for 12 Month Tasks
Purification of Direct Q3 water system
Replace pump oil and Activate Alumina on Roughing Pumps
Maintenance Instructions for 12 Month Tasks
Purification of Direct Q3 water system
Items you will need before beginning:
Smartpak filter (back up on Shelves about KD's Desk)
Millipak 40 filter (back up on Shelves about KD's Desk)
1.2L 30% hydrogen peroxide (Chemical shelving)
Add 1.2 L Hydrogen peroxide to 30L (RO) reservoir tank
Allow to sit 1 hour to sanitize
Run at least 7L through dispenser to sanitize entire system
Can use spigot on RO tank to dispense remaining water
Replace Smartpak and allow tank to fill.
Run at least 7L through dispenser to sanitize entire system
Can use spigot on RO tank to dispense remaining water
Replace Millipak 40 dispenser filter
Reorder Consumables to make sure on hand for next service
Consumables from Fisher Scientific
Item # SPR00SIA1 Millipore Smartpak (water filer cartridge)
Item # MPGL04SK2 Millipore Millipak 40 (dispenser filter)
Item # 382171 4L 30% hydrogen peroxide (need 1.2L)
Replace pump oil and Activate Alumina on Roughing Pumps
Changed:
<
<
* Two pump locations
>
>
Two pump locations
BOC Edwards Auto 306
JEOL film pumping station
Added:
>
>
Items you will need before beginning:
#19 Vacuum pump oil
Activated Alumina
Reorder Consumables to make sure on hand for next service
Consumables from VWR
54996-082 VWR #19 vacuum pump oil (4L)
55009-278 Activated Alumina Charge (500g)