Maintenance Instructions for 12 Month Tasks

Purification of Direct Q3 water system

  • Items you will need before beginning:
    • Smartpak filter (back up on Shelves about KD's Desk)
    • Millipak 40 filter (back up on Shelves about KD's Desk)
    • 1.2L 30% hydrogen peroxide (Chemical shelving)

  • Add 1.2 L Hydrogen peroxide to 30L (RO) reservoir tank
  • Allow to sit 1 hour to sanitize
  • Run at least 7L through dispenser to sanitize entire system
    • Can use spigot on RO tank to dispense remaining water
  • Replace Smartpak and allow tank to fill.
  • Run at least 7L through dispenser to sanitize entire system
    • Can use spigot on RO tank to dispense remaining water
  • Replace Millipak 40 dispenser filter
  • Reorder Consumables to make sure on hand for next service

  • Consumables from Fisher Scientific
    • Item # SPR00SIA1 Millipore Smartpak (water filer cartridge)
    • Item # MPGL04SK2 Millipore Millipak 40 (dispenser filter)
    • Item # 382171 4L 30% hydrogen peroxide (need 1.2L)

Replace pump oil and Activate Alumina on Roughing Pumps

  • Two pump locations
    • BOC Edwards Auto 306
    • JEOL film pumping station

  • Items you will need before beginning:
    • #19 Vacuum pump oil
    • Activated Alumina

  • Reorder Consumables to make sure on hand for next service

  • Consumables from VWR
    • 54996-082 VWR #19 vacuum pump oil (4L)
    • 55009-278 Activated Alumina Charge (500g)
Changed:
<
<

Clean Duct Sox (4)

>
>

Clean Duct Sox (5)

 
  • Turn off air
  • Remove Duct Sox
  • Take to Laundromat
  • Have cleaned
  • install clean sox