Maintenance Instructions for 6 Month Items

F20 PM

  • Call FEI to open service call 1-800-432-1734 Serial #D934

JEOL Microscope PM's

Replace Diaphrams on Gatan Dry Pumping Station

  • Service both FEI and JEOL pumping stations at the same time.
  • You will need:
    • Gatan hex key set from drawers below 2100 pumping station
    • Gatan MVP 015 Service Kit #08503
Added:
>
>
    • 11mm (pump removal) and 14mm (filter removal) crescent wrenches
 
  • Remove Black ring from opening to Turbo pump
  • Remove blue cover
  • Cover opening of turbo pump with tin foil
  • Remove 4 hex screws on both sides of MVP
  • Clean all 6 surfaces of the plates
  • Unscrew diaphragms on each side
    • Make sure new diaphragms are tightly secured or they will strip the arms of the pump requiring a pump replacement.