Maintenance Instructions for 6 Month Items
F20 PM
JEOL Microscope PM's
Replace Diaphrams on Gatan Dry Pumping Station
Maintenance Instructions for 6 Month Items
F20 PM
Call FEI to open service call 1-800-432-1734 Serial #D934
JEOL Microscope PM's
Call or email Tim Gilbert 1-800-253-5365 or
service2@jeol.com
Replace Diaphrams on Gatan Dry Pumping Station
Changed:
<
<
Service both FEI and JEOL pumping stations at the same time.
>
>
Service only pumping station at the same time. Make sure at least one is working.
You will need:
Gatan hex key set from drawers below 2100 pumping station
Gatan MVP 015 Service Kit #08503
11mm (pump removal) and 14mm (filter removal) crescent wrenches
Remove Black ring from opening to Turbo pump
Remove blue cover
Cover opening of turbo pump with tin foil
Remove 4 hex screws on both sides of MVP
Clean all 6 surfaces of the plates
Unscrew diaphragms on each side
Make sure new diaphragms are tightly secured or they will strip the arms of the pump requiring a pump replacement.