Maintenance Instructions for 14 day tasks

Week A

Vacuum F20 and 1230 rooms

Regrease O-rings and clean All Microscope Holders

  • Only Use Fomblin Grease and Lint Free cloths
    • JEOL 1230 (1)
    • JEOL 2100 (3)
    • FEI F20 (5)

Fill U-traps on drains with water

  • 2 locations
    • dark room
    • behind dishwasher

Check Vacuum Gauges on on Gatan Dry Pumping Station

  • This is to ensure that the Cold Cathode Gauge is working
  • Should read 10-4 range if pumped overnight
  • If it does not clean the cold cathode gauge:
    • Turn off and unplug pumping station
    • Vent the system
    • Remove gauge by pulling (it is secured by vacuum and an o-ring)
    • Unscrew the black plastic nut on the bottom of the top gauge.
    • Open gauge.
    • Loosen and remove securing plate with two Hex screws
    • Remove pin
    • With the pin will come : 1 o-ring
    • Use a 3M pad to remove contamination from the pin
      • Use caution and prevent moving the small metal igniter its location is critical to the functioning of the gauge.
    • Dust with compressed air.
    • Reassemble the gauge.

Week B

Vacuum 2100 and 3200 rooms

Check Compressed gas tank regulators

  • Open gas tank
  • Observe pressure on right dial of regulator
  • If at or below 500 psi check end of hallway for back up and list on dry erase board
  • 9 compressed gas tanks
    • HPF
    • Ethane
    • Nitrogen for Plunge Freezer
    • CO2 for incubator
    • F20 Film exchange tank
    • 2100 ancillary hallway
    • 3200 ancillary hallway
    • Gatan Solarius gases (Ar, N, and O)

Check oil levels on all Mechanical pumps

  • Observe level and color of oil in all mechanical pumps
  • If oil shows yellow or darkening hue replace. Do not wait until it browns.
  • 8 Locations - 10 pumps
    • BOC Edwards Vacuum Evaporator
    • JEOL Film pumping station - Microscope Suite Hallway
    • F20 Film pumping station - below phase II stairway
    • F20 Mechanical pump
    • 3200 Mechanical pump (2)
    • 2100 Mechanical pump (2)
    • 1230 Mechanical pump (1)
    • Jun-Aire Compressor -1230 Ancillary hallway

Check Chiller water levels

  • Check 5 locations
    • 4 water chillers - one for each microscope
    • Incubator water level - Cell Culture Room

  • Set ALLOWTOPICVIEW =