
Tests of the micro-dialysis plates performancesReferenceSee MicroDialysisPlatesTool for an overview and information about the protocol.Experiments; test of the plateJune, 23rd 2011 meeting - Tests of the performance of the plateSee ralph_june_08_2011.ppt.
May 2011Before testing leakage and dialysis, it was decided that it is necessary to ensure spectrophotometric assays are reliable. Thus, experiments were carried out to evaluate OD, absorbance, and extinction coefficient measurements. Some of the factors for consideration include: NanoDrop? (Thermo Scientific), UV-Vis spec (Jenway), another UV-Vis spec (Beckman), quartz cuvette, bovine serum albumin (BSA), rose bengal dye (RB), extinction coefficients, dialysis using cassetteApril 2011New plates were obtained from GN Biosystems. Using a solution of Brilliant Blue R dye, leakage thru the microdialysis membrane was investigated. 10 ul of dye solution was placed in the sample chamber of each well. 500 ul of buffer was placed in the buffer chamber of each well. Of the 96 wells, 12 wells did not completely retain the dye in the sample chamber.Sept 2010 (Martin's archive)The overall design of the current plate is fine. The big problem, however, is that the bottom plastic film is very poorly attached to the plate and causes leakage between adjacent buffer chambers.July 2010 (Martin's archive)The performance of the minidialysis plate was evaluated by quantifying the rate of detergent removal through thin layer chromatography (TLC) and by producing crystals of the membrane protein P2A3. For the detergent removal tests, three detergents with very different CMC’s were evaluated: 100mg/ml OG (CMC: 5.3 mg/ml), 30mg/ml DM (CMC: 0.87mg/ml), and 20mg/ml DDM (CMC: 0.087mg/ml). OG was rapidly eliminated from the dialysis wells, and after six buffer exchanges all OG had been removed. Interestingly, the rate of dialysis seems to be a function of time, and independent of buffer exchange frequency (compare figures below). Thus, the gradient difference over the dialysis membrane seems not to be the rate limiting factor for OG removal from the sample well, since frequent buffer exchanges do not increase that rate. In fact, the rate of removal is essentially linear with time for the first 40h of dialysis or until approximately 80% of the detergent has been removed, whereafter the rate slows down considerably. These results indicate that the rate of dialysis may be governed by some intrinsic properties of OG, such as its interaction with the dialysis membrane that may reduce its diffusion speed, or disassembly of the OG micelles.
![]() OG removal with the micro-dialysis plate DM was removed from the dialysis wells at an intermediate rate. After around 10 buffer exchanges or 150h, almost all of the detergent was removed. Note that also for DM the rate seems to be independent of the buffer exchange frequency, which is especially evident for buffer exchange no 7 where the buffer was left in the plate for close to 40h before it was changed, and therefore allowed a lot of DM to be removed at once. Also in this case the there seems to be another rate-limiting step to the speed of removal than the frequency of buffer exchanges, since also here the initial part of the removal curve is essentially linear when plotted against time but not when plotted against buffer exchange number. As expected, DDM was removed considerably slower than the other detergents and required roughly 25 buffer exchanges and a total dialysis time of two weeks. ![]() DDM removal with the micro-dialysis plate These results are similar to the results obtained with the crystallization block (Vink et al., 2007) and confirm that even detergents with very low CMCs can be consistently removed through dialysis. Next, the protein P2A3 (purified in DDM) was dialyzed in the minidialysis plates using conditions previously established with the larger dialysis block and with dialysis buttons. After two weeks of dialysis, with buffer exchanges twice per day, long tubular crystals were observed with a similar abundance to those obtained with the other dialysis devices. These results clearly demonstrate the functionality of the microfluidic dialysis plate and work is currently underway to incorporate the plate into our crystallization pipeline. -- NicolasCoudray - 20 Jan 2011 | |||||||||
| Added: | |||||||||
| > > | * ralph_august25.ppt: cyclodextrin+microdialysis (YiiP?) | ||||||||
| |||||||||
| Added: | |||||||||
| > > |
| ||||||||