New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on the desktop

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 Intensity Zoom
L3 spot size - R3 spot size +

New user Low Dose setup

    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file listed on whiteboard
    • save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties\screen saver\power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\tcl\\EMMENU\\tem1000
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2505 Feb 2008 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

Changed:
<
<
  • Make sure Services is on your desktop Path Programs/Administrative tools/services
>
>
  • Make sure Services is on the desktop
 
  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 Intensity Zoom
L3 spot size - R3 spot size +

New user Low Dose setup

    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file listed on whiteboard
    • save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties\screen saver\power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\tcl\\EMMENU\\tem1000
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2428 Nov 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 Intensity Zoom
L3 spot size - R3 spot size +

New user Low Dose setup

    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file listed on whiteboard
    • save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties\screen saver\power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\TVIPS\\tcl\\EMMENU\\tem1000
>
>
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\tcl\\EMMENU\\tem1000
 
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2323 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
Changed:
<
<
L2 Alpha wobbler R2 screen dim
>
>
L2 Alpha wobbler R2 Intensity Zoom
 
L3 spot size - R3 spot size +

Changed:
<
<

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as... in new user file and folder
>
>

New user Low Dose setup

    • make sure specimen rod is still inserted
      • This will make spot size 2 the default, if rod is out 5 is default.
    • Load known good Low Dose file listed on whiteboard
    • save as new user file and folder
Added:
>
>
 

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties\screen saver\power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\TVIPS\\tcl\\EMMENU\\tem1000
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2208 May 2007 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as... in new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

Changed:
<
<
  • Turn off auto screen off - right click screen: properties/screen saver/power
>
>
  • Turn off auto screen off - right click screen: properties\screen saver\power
 
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
>
>
  • To enable flat correction - File\preferences\misc - change C:\program files\... to C:\\TVIPS\\tcl\\EMMENU\\tem1000
 
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2102 Mar 2007 - Main.BillRice

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
Added:
>
>
    • advanced users should also be Power Users. Instructions for multiple groups:
      • Manage Right click My computer --> Manage --> local Users and Groups --> Users--> (highlight name) --> Properties --> "Member of" tab -->Add...
 
  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as... in new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
  • To place scalebar:
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 2013 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality

  • Remind user to logout and not "shutdown" computers

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds
Changed:
<
<
* save as new user file and folder
>
>
* save as... in new user file and folder
 

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /
Changed:
<
<
>
>
  • To place scalebar:
Added:
>
>
    • scaling/scalebar check "use Scalebar" box
  • 8 bit Tiff
 

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1912 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality
Added:
>
>
  • Remind user to logout and not "shutdown" computers
 

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menuup
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1812 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to never expire
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox

  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Changed:
<
<
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
>
>
Vacuum (expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum (expert) Stigmators (check “hide active control…”)
 
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1712 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
Changed:
<
<
  • Set passwords to not expire
>
>
  • Set passwords to never expire
Added:
>
>
    • Windows 2000 or XP
      • Must be Administrator
      1. Right-click My Computer
      2. Click Manage
      3. Click to Expand Local Users and Groups
      4. Click Users
      5. Double click account
      6. Select Password Never Expires checkbox
 
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Prompted upon opening:
    • User name (first initial and last name)
    • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when prompted upon opening camera
  • Set speed to 300ms
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1612 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Added:
>
>
 
Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +
Deleted:
<
<
 
Added:
>
>
 

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Changed:
<
<
  • Working directory – P: /
>
>
  • Prompted upon opening:
Added:
>
>
    • User name (first initial and last name)
    • Working directory – P: /
 

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

Changed:
<
<
  • Select “firewire” when opening camera
>
>
  • Select “firewire” when prompted upon opening camera
 
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - File/preferences/misc - change C;/ program files to c:/ TVIPS
  • Click flat correction
>
>
  • To enable flat correction - File/preferences/misc - change C:/program files/ to C:/ TVIPS/
  • Check flat correction box
 

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1512 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when opening camera
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - Right click icon: change "start in" path to C:\TVIPS\TCL\Fastscan
>
>
  • To enable flat correction - File/preferences/misc - change C;/ program files to c:/ TVIPS
Deleted:
<
<
    • "target" path C:\TVIPS\TCL\Fastscan\FastScan.exe
 
  • Click flat correction

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1412 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when opening camera
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - Right click icon: change path to C:\TVIPS\TCL\Fastscan
>
>
  • To enable flat correction - Right click icon: change "start in" path to C:\TVIPS\TCL\Fastscan
Added:
>
>
    • "target" path C:\TVIPS\TCL\Fastscan\FastScan.exe
 
  • Click flat correction

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1312 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

  • Working directory – P: /

Fast Scan Camera Computer

  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black

Fast Scan Camera:

  • Select “firewire” when opening camera
  • Set speed to 300ms
Changed:
<
<
  • To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”)
>
>
  • To enable flat correction - Right click icon: change path to C:\TVIPS\TCL\Fastscan
 
  • Click flat correction

Serial EM

  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings

System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing

  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data

  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

  • Focus area- move to bottom of screen so the camera charge will clear.

  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1210 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Changed:
<
<
Working directory – P: /
>
>
  • Working directory – P: /
 

Fast Scan Camera Computer

Changed:
<
<
Turn off auto screen off - right click screen: properties/screen saver/power Make background black
>
>
  • Turn off auto screen off - right click screen: properties/screen saver/power
  • Make background black
 

Fast Scan Camera:

Changed:
<
<
Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction
>
>
  • Select “firewire” when opening camera
  • Set speed to 300ms
  • To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”)
  • Click flat correction
 

Serial EM

Changed:
<
<
Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings
>
>
  • Use C drive
  • Personal directory
  • Settings are local
    • look feel, camera, tilt series
  • Make a shortcut
    • Right click icon to make sure *Start in : Change to user directory * Target is correct
      • Okay to adopt others settings
 
Changed:
<
<
System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing
>
>
System Vs User directories
Program personal settings
Calibrations need to change start in line under properties to prevent over writing
 
Changed:
<
<
Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories
C
Technai Data
>
>
  • Load alignments for F20 and assume the calibrations for serial EM are valid.
    • Stress that it is a bad idea to fiddle with alignments
    • FEG registers Alignments
    • User directories
    • C: Technai Data
 
Changed:
<
<
Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.
>
>
  • Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.
 
Changed:
<
<
Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.
>
>
  • Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.
 
Changed:
<
<
Focus area- move to bottom of screen so the camera charge will clear.
>
>
  • Focus area- move to bottom of screen so the camera charge will clear.
 
Changed:
<
<
16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.
>
>
  • 16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.
 
Changed:
<
<
Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
>
>
  • Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view.
 Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1110 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Changed:
<
<
* TUIpanel.bmp: TUIpanel.bmp
>
>
Deleted:
<
<
 
META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 1010 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +
Added:
>
>
 
Added:
>
>
 

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006 * TUIpanel.bmp: TUIpanel.bmp

Deleted:
<
<
* TUIpanel.bmp: TUIpanel.bmp
 
META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 910 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Added:
>
>
* TUIpanel.bmp: TUIpanel.bmp
  * TUIpanel.bmp: TUIpanel.bmp

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 810 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006 * TUIpanel.bmp: TUIpanel.bmp

META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 710 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Added:
>
>
* TUIpanel.bmp: TUIpanel.bmp
 
META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"

Revision 610 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Added:
>
>
META FILEATTACHMENT attachment="TUIpanel.bmp" attr="" comment="" date="1160502976" name="TUIpanel.bmp" path="TUIpanel.bmp" size="712856" stream="TUIpanel.bmp" user="Main.KdDerr" version="1"
 

Revision 510 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +
Added:
>
>

Low Dose setup

* make sure specimen rod is inserted to first stop * This will make spot size 2 the default, if rod is out spot size is clipped to 5. * Load known good Low Dose file like KD0.lds * save as new user file and folder
 

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Revision 405 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Changed:
<
<
Vacuum Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Setup
High tension Plate (set date and path to record image info in users folder) FEG Controls Direct alignments Vacuum Stigmators (check “hide active control…”)
  Exposure log     Beam Settings
>
>
Vacuum ( expert) Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Stage 2
High tension Plate Camera (user) (set date and path to record image info in users folder) Direct alignments Vacuum Stigmators (check “hide active control…”)
FEG control (expert) Exposure log      
 

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Revision 305 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads
Changed:
<
<

TECNAI USER INTERFACE TABS |

>
>

TECNAI USER INTERFACE TABS

 
Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Setup
High tension Plate (set date and path to record image info in users folder) FEG Controls Direct alignments Vacuum Stigmators (check “hide active control…”)
  Exposure log     Beam Settings

TECNAI USER INTERFACE LOWER SCREEN PANELS

Changed:
<
<
Button designations L1 resets defocus R1 screen lift L2 Alpha wobbler R2 screen dim L3 spot size - R3 spot size +
>
>
Button designations
L1 resets defocus R1 screen lift
L2 Alpha wobbler R2 screen dim
L3 spot size - R3 spot size +
 

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Revision 205 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads
Changed:
<
<

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Vacuum Detector Configuration (uncheck “Auto Mode”) High tension Plate (set date and path to record image info in users folder) FEG Controls Exposure log
>
>

TECNAI USER INTERFACE TABS |

Setup menu Camera Menu Tune Menu Low Dose Search Menu
Vacuum Detector Configuration (uncheck “Auto Mode”) Alignments Low dose Setup
High tension Plate (set date and path to record image info in users folder) FEG Controls Direct alignments Vacuum Stigmators (check “hide active control…”)
Added:
>
>
  Exposure log     Beam Settings
 
Deleted:
<
<
Tune Menu Low Dose Search Menu Alignments Low dose Setup Direct alignments Vacuum Stigmators (check “hide active control…”) Beam Settings
 

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations L1 resets defocus R1 screen lift L2 Alpha wobbler R2 screen dim L3 spot size - R3 spot size +

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.

  • Set ALLOWTOPICVIEW =

-- KdDerr - 04 Oct 2006

Revision 104 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CemAdmin"
Contents

New User Setup Tecnai F20

  • Create new accounts on both computers and Greymatter
  • Set passwords to not expire
  • Set up new users as “TEMexperts” to insure functionality

Tecnai Computer Settings

  • Make sure Services is on your desktop Path Programs/Administrative tools/services

  • Tecnai User Interface C:/Tecnai/exe/Peoui.exe

  • Put control pad light switch on menu bar – tools/customize tool bar/commands/control pads

TECNAI USER INTERFACE TABS

Setup menu Camera Menu Vacuum Detector Configuration (uncheck “Auto Mode”) High tension Plate (set date and path to record image info in users folder) FEG Controls Exposure log

Tune Menu Low Dose Search Menu Alignments Low dose Setup Direct alignments Vacuum Stigmators (check “hide active control…”) Beam Settings

TECNAI USER INTERFACE LOWER SCREEN PANELS

Button designations L1 resets defocus R1 screen lift L2 Alpha wobbler R2 screen dim L3 spot size - R3 spot size +

Teitz Software

Working directory – P: /

Fast Scan Camera Computer

Turn off auto screen off - right click screen: properties/screen saver/power Make background black

Fast Scan Camera:

Select “firewire” when opening camera Set speed to 300ms To enable flat correction - Right click icon: change path to C: /TVIPS/…. (replace “program files”) Click flat correction

Serial EM

Use C drive Personal directory Settings are local -look feel, camera, tilt series Make a shortcut Right click icon to make sure Start in : Change to user directory Target is correct Okay to adopt others settings

System Vs User directories Program personal settings Calibrations need to change start in line under properties to prevent over writing

Load alignments for F20 and assume the calibrations for serial EM are valid. Stress that it is a bad idea to fiddle with alignments FEG registers Alignments User directories

C
Technai Data

Calibrations must be done in admin mode. To save calibrations you must check next to admin in pull down menu. DO not calibrate unless authorized.

Binning- adds up selected number of pixels and makes them 1. Reduces info. Shortens exposure.

Focus area- move to bottom of screen so the camera charge will clear.

16 Bits- camera reads 0-65000. Prefer to truncate data, anything over 32500 is thrown away.

Low Dose- when Tecnai is in exposure mode open serial EM low dose and check box. Focus and tracking should be in the same place. The beam in these areas should not overlap record area. Do not normalize beam through view. Start with a negative tilt and go positive. Nudge goniometer up higher to avoid hysterisis.