lineshape sample the non-spin numbers are 5.8/8.2/0.57, however the satellite peaks appear weak, 0.1%EB samples the best 200Hz s:n we get is 1745:1 (should be over 2000:1, Water suppression on 2mM sucrose: resolution is 15%, s:n is 364:1, and water suppression is 42/69. Brian has scheduled Jeff Parent to come in of 6/7 to evaluate the probe.
Poor 1H sensitivity
OPEN
5/26/05
Emailed Brian Lord at Bruker
2mM Sucrose S:N ~312:1 down from 450:1/ We will make further attempts to shim the probe in.
Poor 1H sensitivity
CLOSED
6/7/05
Jeff Parent on-site to evaluate probe.
Jeff was able to get lineshape to 4.7/11.1/ , 0.1%EB sensitivity numbers for 200Hz s:n is 2300:1 and water suppresion are, resolution 22%, s:n 407:1 and hump 44.8/107.2