Difference: PowerMeasurements600 (1 vs. 4)

Revision 424 Oct 2006 - Main.ShibaniBhattacharya

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 600

Low Power

Low power measurements: As of 12/8/05

The low power measurements were done with 1GHz oscilloscope via 1m RG241 cable. The high power measurements were done with power meter via approx. 40dB attn. The data in the tables is corrected for attenuation. 15N and 13C amplifier outputs were non-linear between -6dB and -3dB

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 924mV         1.04V
1H   848mV       1.04V
15N     988mV   1.33V 1.67V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High Power

Added:
>
>
High power measurements: As of 10/24/06 (After Replacing 1H/13C Amplifier)

Measurement description BLAX300 (13C), W BLAH (1H), W
amplifier output @ -6dB 0 dB attenuation 326/358 64.3/70.7
amplifier output @ -6dB w/ attenuation 214/235 (4 dB) 214/235 (9 dB)
preamp output @ -6dB w/ attenuation 139/153 (4 dB) 7.24/7.94 (9 dB)
probe input @ -5dB w/ attenuation 111/122 (4 dB)
Pulse Lengths @ -5dB w/ attenuation 13us (4 dB) 8.25us (9 dB)
 High power measurements: As of 10/18/06 (Before removing all inline attenuators)

13C 90 Pulse length increased from 18 us @ -6.0 dB to 21.5 us @ -6 dB. Could see phase distortions in spectra. Removed 2 dB inline attenuator to get 18 us @ -6 dB and recortabed the amplifier output.

Changed:
<
<
Measurement description BLAX300 (13C), W
>
>
Measurement description BLAX (13C), W
 
amplifier output @ -6dB 0 dB attenuation 75.8/84.2
probe input @ -6dB with 0 dB attenuation 49.3/54.8
probe input @ -6dB with 2 dB attenuation 34.9/38.9

High power measurements: As of 02/22/06 (After replacing 15N Amplifier)

Measurement description BLAX300 (15N), W
amplifier output @ -6dB no attenuation 461/489
amplifier output @ -6dB with attenuation 2dB, 433/459
probe input @ -6dB with attenuation 277/294
probe input @ -4.5dB with attenuation 193/205

40 us 15N 90 @ -4.3 dB (Probe spec'd for 222 W)

High power measurements made on 12/08/05 by Shibani: 600MHz_power_measurements.doc:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W
amplifier output @ -6dB no attenuation 191/217.7 51.5/68.5 311/335.9
amplifier output @ -6dB with attenuation 4dB, 98.2/112 9dB, 12.3/13.16 2dB, 241/260.3
probe input @ -6dB with attenuation 64.3/73.3 7.02/7.51 151/163.1

Linearity Check

Pulse calibration done on Urea/Methanol in DMSO

Nucleus Power level, dB Pulse length, us, 05/25/05 Pulse length, us, 05/17/05 Pulse length, us, 04/24/05
C13 -5dB 15 16.5 14
  1 30    
  7 63    
1H -5 8.12    
  -3 10.19    
  0 14.25    
  3 20.5    
  6 28.25    
15N -6 40 40  
  -5 44 49 43
  0 84    
  6 170    
  7.3 190    
2H non linear ???
  - DMSO Sucrose (90%H2O/10%D2O)  
  -6 116.5 -  
  -3 165.0 -  
  0 212.5 211.5  
  3.0 - -    
  5.2 - 380  
  6.0 410 -  



Previous Measurements

Low power measurements:

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 916/934mV         1.02/1.04V
1H   850/901mV       932/987mV
15N     996V   1.37V 1.65V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High power measurements:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W BLAX3 (2H), W
amplifier output @ -6dB no attenuation 268/300 61/80 390/413 380/418
amplifier output @ -6dB with attenuation 4dB, 153/171 9dB, 11.5/15 2dB, 333/352 10dB, 36.6/40
probe input @ -6dB with attenuation 100/112 7.5/9.75 215/228 10/10.6

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 14 Dec 2005

META FILEATTACHMENT attr="" comment="" date="1134572232" name="600MHz_power_measurements.doc" path="600 MHz_power_measurements.doc" size="42496" user="MichaelGoger" version="1.1"

Revision 318 Oct 2006 - Main.ShibaniBhattacharya

 
META TOPICPARENT name="NmrStaffnotes"
Changed:
<
<
Contents

Power Measurements for 600

Low Power

Low power measurements: As of 12/8/05

The low power measurements were done with 1GHz oscilloscope via 1m RG241 cable. The high power measurements were done with power meter via approx. 40dB attn. The data in the tables is corrected for attenuation. 15N and 13C amplifier outputs were non-linear between -6dB and -3dB

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 924mV         1.04V
1H   848mV       1.04V
15N     988mV   1.33V 1.67V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High Power

High power measurements: As of 02/22/06 (After replacing 15N Amplifier)

Measurement description BLAX300 (15N), W
amplifier output @ -6dB no attenuation 461/489
amplifier output @ -6dB with attenuation 2dB, 433/459
probe input @ -6dB with attenuation 277/294
probe input @ -4.5dB with attenuation 193/205

40 us 15N 90 @ -4.3 dB (Probe spec'd for 222 W)

High power measurements made on 12/08/05 by Shibani: 600MHz_power_measurements.doc:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W
amplifier output @ -6dB no attenuation 191/217.7 51.5/68.5 311/335.9
amplifier output @ -6dB with attenuation 4dB, 98.2/112 9dB, 12.3/13.16 2dB, 241/260.3
probe input @ -6dB with attenuation 64.3/73.3 7.02/7.51 151/163.1

Linearity Check

Pulse calibration done on Urea/Methanol in DMSO

Nucleus Power level, dB Pulse length, us, 05/25/05 Pulse length, us, 05/17/05 Pulse length, us, 04/24/05
C13 -5dB 15 16.5 14
  1 30    
  7 63    
1H -5 8.12    
  -3 10.19    
  0 14.25    
  3 20.5    
  6 28.25    
15N -6 40 40  
  -5 44 49 43
  0 84    
  6 170    
  7.3 190    
2H non linear ???
  - DMSO Sucrose (90%H2O/10%D2O)  
  -6 116.5 -  
  -3 165.0 -  
  0 212.5 211.5  
  3.0 - -    
  5.2 - 380  
  6.0 410 -  



Previous Measurements

Low power measurements:

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 916/934mV         1.02/1.04V
1H   850/901mV       932/987mV
15N     996V   1.37V 1.65V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High power measurements:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W BLAX3 (2H), W
amplifier output @ -6dB no attenuation 268/300 61/80 390/413 380/418
amplifier output @ -6dB with attenuation 4dB, 153/171 9dB, 11.5/15 2dB, 333/352 10dB, 36.6/40
probe input @ -6dB with attenuation 100/112 7.5/9.75 215/228 10/10.6

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 14 Dec 2005

>
>
Contents

Power Measurements for 600

Low Power

Low power measurements: As of 12/8/05

The low power measurements were done with 1GHz oscilloscope via 1m RG241 cable. The high power measurements were done with power meter via approx. 40dB attn. The data in the tables is corrected for attenuation. 15N and 13C amplifier outputs were non-linear between -6dB and -3dB

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 924mV         1.04V
1H   848mV       1.04V
15N     988mV   1.33V 1.67V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High Power

High power measurements: As of 10/18/06 (Before removing all inline attenuators)

13C 90 Pulse length increased from 18 us @ -6.0 dB to 21.5 us @ -6 dB. Could see phase distortions in spectra. Removed 2 dB inline attenuator to get 18 us @ -6 dB and recortabed the amplifier output.

Measurement description BLAX300 (13C), W
amplifier output @ -6dB 0 dB attenuation 75.8/84.2
probe input @ -6dB with 0 dB attenuation 49.3/54.8
probe input @ -6dB with 2 dB attenuation 34.9/38.9

High power measurements: As of 02/22/06 (After replacing 15N Amplifier)

Measurement description BLAX300 (15N), W
amplifier output @ -6dB no attenuation 461/489
amplifier output @ -6dB with attenuation 2dB, 433/459
probe input @ -6dB with attenuation 277/294
probe input @ -4.5dB with attenuation 193/205

40 us 15N 90 @ -4.3 dB (Probe spec'd for 222 W)

High power measurements made on 12/08/05 by Shibani: 600MHz_power_measurements.doc:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W
amplifier output @ -6dB no attenuation 191/217.7 51.5/68.5 311/335.9
amplifier output @ -6dB with attenuation 4dB, 98.2/112 9dB, 12.3/13.16 2dB, 241/260.3
probe input @ -6dB with attenuation 64.3/73.3 7.02/7.51 151/163.1

Linearity Check

Pulse calibration done on Urea/Methanol in DMSO

Nucleus Power level, dB Pulse length, us, 05/25/05 Pulse length, us, 05/17/05 Pulse length, us, 04/24/05
C13 -5dB 15 16.5 14
  1 30    
  7 63    
1H -5 8.12    
  -3 10.19    
  0 14.25    
  3 20.5    
  6 28.25    
15N -6 40 40  
  -5 44 49 43
  0 84    
  6 170    
  7.3 190    
2H non linear ???
  - DMSO Sucrose (90%H2O/10%D2O)  
  -6 116.5 -  
  -3 165.0 -  
  0 212.5 211.5  
  3.0 - -    
  5.2 - 380  
  6.0 410 -  



Previous Measurements

Low power measurements:

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 916/934mV         1.02/1.04V
1H   850/901mV       932/987mV
15N     996V   1.37V 1.65V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High power measurements:

Added:
>
>
Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W BLAX3 (2H), W
amplifier output @ -6dB no attenuation 268/300 61/80 390/413 380/418
amplifier output @ -6dB with attenuation 4dB, 153/171 9dB, 11.5/15 2dB, 333/352 10dB, 36.6/40
probe input @ -6dB with attenuation 100/112 7.5/9.75 215/228 10/10.6

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 14 Dec 2005

 
META FILEATTACHMENT attr="" comment="" date="1134572232" name="600MHz_power_measurements.doc" path="600 MHz_power_measurements.doc" size="42496" user="MichaelGoger" version="1.1"

Revision 223 Feb 2006 - Main.ShibaniBhattacharya

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 600

Low Power

Low power measurements: As of 12/8/05

The low power measurements were done with 1GHz oscilloscope via 1m RG241 cable. The high power measurements were done with power meter via approx. 40dB attn. The data in the tables is corrected for attenuation. 15N and 13C amplifier outputs were non-linear between -6dB and -3dB

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 924mV         1.04V
1H   848mV       1.04V
15N     988mV   1.33V 1.67V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High Power

Added:
>
>
High power measurements: As of 02/22/06 (After replacing 15N Amplifier)

Measurement description BLAX300 (15N), W
amplifier output @ -6dB no attenuation 461/489
amplifier output @ -6dB with attenuation 2dB, 433/459
probe input @ -6dB with attenuation 277/294
probe input @ -4.5dB with attenuation 193/205

40 us 15N 90 @ -4.3 dB (Probe spec'd for 222 W)

 High power measurements made on 12/08/05 by Shibani: 600MHz_power_measurements.doc:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W
amplifier output @ -6dB no attenuation 191/217.7 51.5/68.5 311/335.9
amplifier output @ -6dB with attenuation 4dB, 98.2/112 9dB, 12.3/13.16 2dB, 241/260.3
probe input @ -6dB with attenuation 64.3/73.3 7.02/7.51 151/163.1

Linearity Check

Pulse calibration done on Urea/Methanol in DMSO

Nucleus Power level, dB Pulse length, us, 05/25/05 Pulse length, us, 05/17/05 Pulse length, us, 04/24/05
C13 -5dB 15 16.5 14
  1 30    
  7 63    
1H -5 8.12    
  -3 10.19    
  0 14.25    
  3 20.5    
  6 28.25    
15N -6 40 40  
  -5 44 49 43
  0 84    
  6 170    
  7.3 190    
2H non linear ???
  - DMSO Sucrose (90%H2O/10%D2O)  
  -6 116.5 -  
  -3 165.0 -  
  0 212.5 211.5  
  3.0 - -    
  5.2 - 380  
  6.0 410 -  



Previous Measurements

Low power measurements:

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 916/934mV         1.02/1.04V
1H   850/901mV       932/987mV
15N     996V   1.37V 1.65V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High power measurements:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W BLAX3 (2H), W
amplifier output @ -6dB no attenuation 268/300 61/80 390/413 380/418
amplifier output @ -6dB with attenuation 4dB, 153/171 9dB, 11.5/15 2dB, 333/352 10dB, 36.6/40
probe input @ -6dB with attenuation 100/112 7.5/9.75 215/228 10/10.6

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 14 Dec 2005

META FILEATTACHMENT attr="" comment="" date="1134572232" name="600MHz_power_measurements.doc" path="600 MHz_power_measurements.doc" size="42496" user="MichaelGoger" version="1.1"

Revision 114 Dec 2005 - Main.MichaelGoger

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 600

Low Power

Low power measurements: As of 12/8/05

The low power measurements were done with 1GHz oscilloscope via 1m RG241 cable. The high power measurements were done with power meter via approx. 40dB attn. The data in the tables is corrected for attenuation. 15N and 13C amplifier outputs were non-linear between -6dB and -3dB

Nucleusup SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
2H       960/980mV 1.09/1.11V 1.09/1.11V
1H   848mV       1.04V
15N     988mV   1.33V 1.67V
13C 924mV         1.04V

High Power

High power measurements made on 12/08/05 by Shibani: 600MHz_power_measurements.doc:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W
amplifier output @ -6dB no attenuation 191/217.7 51.5/68.5 311/335.9
amplifier output @ -6dB with attenuation 4dB, 98.2/112 9dB, 12.3/13.16 2dB, 241/260.3
probe input @ -6dB with attenuation 64.3/73.3 7.02/7.51 151/163.1

Linearity Check

Pulse calibration done on Urea/Methanol in DMSO

Nucleus Power level, dB Pulse length, us, 05/25/05 Pulse length, us, 05/17/05 Pulse length, us, 04/24/05
C13 -5dB 15 16.5 14
  1 30    
  7 63    
1H -5 8.12    
  -3 10.19    
  0 14.25    
  3 20.5    
  6 28.25    
15N -6 40 40  
  -5 44 49 43
  0 84    
  6 170    
  7.3 190    
2H non linear ???
  - DMSO Sucrose (90%H2O/10%D2O)  
  -6 116.5 -  
  -3 165.0 -  
  0 212.5 211.5  
  3.0 - -    
  5.2 - 380  
  6.0 410 -  



Previous Measurements

Low power measurements:

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 916/934mV         1.02/1.04V
1H   850/901mV       932/987mV
15N     996V   1.37V 1.65V
2H       960/980mV 1.09/1.11V 1.09/1.11V

High power measurements:

Measurement description BLAX1 (13C), W BLAH (1H), W BLAX2 (15N), W BLAX3 (2H), W
amplifier output @ -6dB no attenuation 268/300 61/80 390/413 380/418
amplifier output @ -6dB with attenuation 4dB, 153/171 9dB, 11.5/15 2dB, 333/352 10dB, 36.6/40
probe input @ -6dB with attenuation 100/112 7.5/9.75 215/228 10/10.6

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 14 Dec 2005

META FILEATTACHMENT attr="" comment="" date="1134572232" name="600MHz_power_measurements.doc" path="600 MHz_power_measurements.doc" size="42496" user="MichaelGoger" version="1.1"
 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.