Difference: PowerMeasurements800us (1 vs. 14)

Revision 1425 May 2007 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#2

Low Power

Low power measurements: As of

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA

High power measurements: As of 5/22/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation -
-
494
-
probe input @ -6dB no attenuation -
-
326
-
amplifier output @ -6dB with attenuation 8dB 234 7dB 17.7 5dB 280 22dB 2.22
probe input @ -6dB with attenuation 8dB 146
7dB 9.75 5dB 175 22dB 0.59
linearity at probe with attn -6dB to +6dB 146/75.3/37.9/19.2/9.82
9.75/5.01/2.47/1.23/0.61
175/81.9/39.7/20.2/9.94
0.59/0.28/0.14/0.07/0.03
linearity at amplifier with attn -6dB to +6dB 234/122/60.6/30.4/15.5 17.7/9.06/4.43/2.22/1.09
280/134/66.2/33.8/16.7
2.22/1.04/0.52/0.25/0.13
Changed:
<
<
linearity at probe without attn -6dB to +6dB -
-
326/115/30.3

linearity at amplifier without attn -6dB to +6dB

494/349/190/102/51.0

>
>
linearity at probe without attn -6dB to +6dB - - 326/115/30.3 -
linearity at amplifier without attn -6dB to +6dB - - 494/349/190/102/51.0 -
 

High Power

High power measurements: As of 2/21/07
Measurement description BLAH100(1H), W BLAX1500 (13C), Wup BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 7dB 15.4/16.9 8dB 256/271 1dB 278/286
pulse lengths 8.05@-5.6
15@-6
44@-6

High power measurements: As of 1/12/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 1325 May 2007 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#2

Low Power

>
>

Power Measurements for 800us#2

Low Power

 Low power measurements: As of

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA

High power measurements: As of 5/22/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
Changed:
<
<
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

>
>
amplifier output @ -6dB no attenuation -
-
494
-
probe input @ -6dB no attenuation -
-
326
-
amplifier output @ -6dB with attenuation 8dB 234 7dB 17.7 5dB 280 22dB 2.22
probe input @ -6dB with attenuation 8dB 146
7dB 9.75 5dB 175 22dB 0.59
linearity at probe with attn -6dB to +6dB 146/75.3/37.9/19.2/9.82
9.75/5.01/2.47/1.23/0.61
175/81.9/39.7/20.2/9.94
0.59/0.28/0.14/0.07/0.03
linearity at amplifier with attn -6dB to +6dB 234/122/60.6/30.4/15.5 17.7/9.06/4.43/2.22/1.09
280/134/66.2/33.8/16.7
2.22/1.04/0.52/0.25/0.13
Added:
>
>
linearity at probe without attn -6dB to +6dB -
-
326/115/30.3

linearity at amplifier without attn -6dB to +6dB

494/349/190/102/51.0

 
Changed:
<
<
>
>

High Power

Deleted:
<
<

High Power

 High power measurements: As of 2/21/07
Measurement description BLAH100(1H), W BLAX1500 (13C), W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 7dB 15.4/16.9 8dB 256/271 1dB 278/286
pulse lengths 8.05@-5.6
15@-6
44@-6

High power measurements: As of 1/12/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 1225 May 2007 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#2

Low Power

Low power measurements: As of

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA
Added:
>
>
High power measurements: As of 5/22/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

 

High Power

High power measurements: As of 2/21/07
Measurement description BLAH100(1H), W BLAX1500 (13C), W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 7dB 15.4/16.9 8dB 256/271 1dB 278/286
pulse lengths 8.05@-5.6
15@-6
44@-6

High power measurements: As of 1/12/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 1121 Feb 2007 - Main.ShibaniBhattacharya

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#2

Low Power

Low power measurements: As of

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA

High Power

Added:
>
>
High power measurements: As of 2/21/07
Measurement description BLAH100(1H), W BLAX1500 (13C), W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 7dB 15.4/16.9 8dB 256/271 1dB 278/286
pulse lengths 8.05@-5.6
15@-6
44@-6
 High power measurements: As of 1/12/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 1002 Feb 2007 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#2

Low Power

>
>

Power Measurements for 800us#2

Low Power

 Low power measurements: As of

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA
Changed:
<
<

High Power

>
>

High Power

 High power measurements: As of 1/12/07 (without correction)

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
Changed:
<
<
linearity at amplifier with attn -6dB to +6dB 131.0; 27.9; 6.4
10.2; 2.24; 0.51
351;86;21.2

>
>
linearity at amplifier with attn -6dB to +6dB 212;44.0;10
10.2; 2.24; 0.51
351;86;21.2

 
linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

Deleted:
<
<
 
  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 902 Feb 2007 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#1

Low Power

>
>

Power Measurements for 800us#2

Low Power

 
Changed:
<
<
Low power measurements: As of 6/08/05
>
>
Low power measurements: As of
 
Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
Changed:
<
<
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13
>
>
13C NA
NA
1H NA
NA
15N NA
NA
NA
2H NA
NA
NA
 
Changed:
<
<

High Power

High power measurements: As of 6/08/05
>
>

High Power

High power measurements: As of 1/12/07 (without correction)
 
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
Changed:
<
<
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
>
>
amplifier output @ -6dB no attenuation 555
72.7
399

probe input @ -6dB no attenuation 331
38.4
248
amplifier output @ -6dB with attenuation 9dB 212 8dB 10.2 1dB 351
probe input @ -6dB with attenuation 9dB 131
8dB 5.74 1dB 216
linearity at amplifier with attn -6dB to +6dB 131.0; 27.9; 6.4
10.2; 2.24; 0.51
351;86;21.2

Added:
>
>
linearity at amplifier without attn -6dB to +6dB 555;473; 316; 181;93.5
72.7;45.3;24.0;11.6;5.21 399;207;104;53.2;25.7

 
Deleted:
<
<
High power measurements: As of 12/05/06
 
Deleted:
<
<
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39 9.45; 2.35; 0.58 308.0; 70.9; 17.8 12.0; 3.04; 0.76


Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB

Power measurement of NEW 13C Amplifier 5/30/06 and 12/05/06

13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0

13C db @Amplifier 1 @probe (W) 9db attenuation 5/30/06 @probe (W) 8db attenuation 12/05/06
-6 - 127.0 127.0
-3 - 65.0 -
0 - 33.5 31.4
3 - 16.2 -
6 - 8.3 8.13

13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -
 
  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 811 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#1

Low Power

>
>

Power Measurements for 800us#1

Low Power

 Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13
Changed:
<
<

High Power

>
>

High Power

 High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46

High power measurements: As of 12/05/06

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
Changed:
<
<
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39 9.45; 2.35; 0.58 308.0; 70.9; 17.8 12.0; 3.04; 0.76
>
>
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39 9.45; 2.35; 0.58 308.0; 70.9; 17.8 12.0; 3.04; 0.76
 

Changed:
<
<

Previous measurements from 3/24/05

>
>

Previous measurements from 3/24/05

 
1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
Changed:
<
<

Power measurement of NEW 13C Amplifier 5/30/06 and 12/05/06

>
>

Power measurement of NEW 13C Amplifier 5/30/06 and 12/05/06

 
13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0

13C db @Amplifier 1 @probe (W) 9db attenuation 5/30/06 @probe (W) 8db attenuation 12/05/06
-6 - 127.0 127.0
-3 - 65.0 -
0 - 33.5 31.4
Changed:
<
<
3 - 16.2 -
>
>
3 - 16.2 -
 
6 - 8.3 8.13

13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 711 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
Changed:
<
<
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
>
>
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
 High power measurements: As of 12/05/06

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
Changed:
<
<
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39
9.45; 2.35; 0.58
308.0; 70.9; 17.8
12.0; 3.04; 0.76
>
>
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39 9.45; 2.35; 0.58 308.0; 70.9; 17.8 12.0; 3.04; 0.76
 

Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
Changed:
<
<
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50
>
>
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50
 
13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
Changed:
<
<
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3
>
>
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3
 
15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB

Power measurement of NEW 13C Amplifier 5/30/06 and 12/05/06

13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0
Changed:
<
<
13C db @Amplifier 1 @probe (W) 9db attenuation After Cortab 5/30/06 @probe (W) 8db attenuation After Cortab 12/05/06
-6 - 127.0 127.0
-3 - 65.0 -
0 - 33.5 31.4
3 - 16.2
6 - 8.3 8.13
>
>
13C db @Amplifier 1 @probe (W) 9db attenuation 5/30/06 @probe (W) 8db attenuation 12/05/06
-6 - 127.0 127.0
-3 - 65.0 -
0 - 33.5 31.4
3 - 16.2 -
6 - 8.3 8.13
 
13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 611 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#1

Low Power

>
>

Power Measurements for 800us#1

Low Power

 Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13
Changed:
<
<

High Power

>
>

High Power

 High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46

High power measurements: As of 12/05/06

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39
9.45; 2.35; 0.58
308.0; 70.9; 17.8
12.0; 3.04; 0.76


Changed:
<
<

Previous measurements from 3/24/05

>
>

Previous measurements from 3/24/05

 
1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
Changed:
<
<

Power measurement of NEW 13C Amplifier 5/30/06

>
>

Power measurement of NEW 13C Amplifier 5/30/06 and 12/05/06

 
13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0
Changed:
<
<
13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3
>
>
13C db @Amplifier 1 @probe (W) 9db attenuation After Cortab 5/30/06 @probe (W) 8db attenuation After Cortab 12/05/06
-6 - 127.0 127.0
-3 - 65.0 -
0 - 33.5 31.4
3 - 16.2
6 - 8.3 8.13
 
13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 511 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46

High power measurements: As of 12/05/06

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
Deleted:
<
<
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
 
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39
9.45; 2.35; 0.58
308.0; 70.9; 17.8
12.0; 3.04; 0.76


Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB

Power measurement of NEW 13C Amplifier 5/30/06

13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0

13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3

13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 411 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents
Changed:
<
<

Power Measurements for 800us#1

Low Power

>
>

Power Measurements for 800us#1

Low Power

 Low power measurements: As of 6/08/05
Changed:
<
<
Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026         1.08/1.14V
1H   830/954mV       912/1.05V
15N     1V/1.04   1.32/1.37 1.6/1.66V
2H       988/1.04mV 1.11/1.16V 1.08/1.13
>
>
Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026 1.08/1.14V
1H 830/954mV 912/1.05V
15N 1V/1.04 1.32/1.37 1.6/1.66V
2H 988/1.04mV 1.11/1.16V 1.08/1.13
 
Changed:
<
<

High Power

>
>

High Power

 High power measurements: As of 6/08/05
Changed:
<
<
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
>
>
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
 
Deleted:
<
<
 High power measurements: As of 12/05/06
Changed:
<
<
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
>
>
Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 8 dB 127/146.7 13dB 8.67/9.45 0 dB 267/308 7 dB 10.8/12.0
linearity at amplifier with attn -6dB to +6dB 146.7;36.3;9.39
9.45; 2.35; 0.58
308.0; 70.9; 17.8
12.0; 3.04; 0.76
 

Changed:
<
<

Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50
>
>

Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50
 
Changed:
<
<
13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3
>
>
13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3
 
Changed:
<
<
15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB
>
>
15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB
 
Changed:
<
<
2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
>
>
2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
 
Changed:
<
<

Power measurement of NEW 13C Amplifier 5/30/06

>
>

Power measurement of NEW 13C Amplifier 5/30/06

 
Added:
>
>
13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0
 
Changed:
<
<
13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0
>
>
13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3
 
Changed:
<
<
13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3
>
>
13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -
 
Changed:
<
<
13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
>
>
  • Set ALLOWTOPICVIEW =
Deleted:
<
<
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -
 
Changed:
<
<
>
>
-- MichaelGoger - 13 Dec 2005
Deleted:
<
<
  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 311 Dec 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026         1.08/1.14V
1H   830/954mV       912/1.05V
15N     1V/1.04   1.32/1.37 1.6/1.66V
2H       988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
Added:
>
>
High power measurements: As of 12/05/06

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46
 

Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB

Power measurement of NEW 13C Amplifier 5/30/06

13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0

13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3

13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 231 May 2006 - Main.KaushikDutta

 
META TOPICPARENT name="NmrStaffnotes"
Changed:
<
<
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026         1.08/1.14V
1H   830/954mV       912/1.05V
15N     1V/1.04   1.32/1.37 1.6/1.66V
2H       988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46



Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

>
>
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026         1.08/1.14V
1H   830/954mV       912/1.05V
15N     1V/1.04   1.32/1.37 1.6/1.66V
2H       988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46



Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB

Power measurement of NEW 13C Amplifier 5/30/06

Added:
>
>
13C db @Amplifier 1 @Amp NO attenution Before Cortab
-6 - 538.0
-3 - 465.0
0 - 335.0
3 - 216.0
6 - 124.0

13C db @Amplifier 1 @probe (W) 9db attenution After Cortab
-6 - 127.0
-3 - 65.0
0 - 33.5
3 - 16.2
6 - 8.3

13C db @Amplifier 1 @Amp (W) 9db attenution After Cortab
-6 - 218.0
-3 - -
0 - -
3 - -
6 - -

  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

Revision 113 Dec 2005 - Main.MichaelGoger

 
META TOPICPARENT name="NmrStaffnotes"
Contents

Power Measurements for 800us#1

Low Power

Low power measurements: As of 6/08/05

Nucleus SGU1 output SGU2 output SGU3 output SGU4 output Router 2 output Amplifier input
13C 968/1026         1.08/1.14V
1H   830/954mV       912/1.05V
15N     1V/1.04   1.32/1.37 1.6/1.66V
2H       988/1.04mV 1.11/1.16V 1.08/1.13

High Power

High power measurements: As of 6/08/05

Measurement description BLAX1 500(13C), W BLAH100, W BLAX2 300 (15N), W BLAX3 300(2H), W
amplifier output @ -6dB no attenuation 482/554 80/88 452/497 320/360
probe input @ -6dB no attenuation 283/325 40/44 283/311 77/86
amplifier output @ -6dB with attenuation 8dB 193/222 13dB 13.7/14.9 1dB 434/477 7dB 85/96
probe input @ -6dB with attenuation 112/129 7/7.7 271/298 20.2/22.8
linearity at amplifier with attn -6dB to +6dB 19.3 9.6 4.7 2.34 1.18 13.7 7.6 3.35 1.70 0.81 43.5 20.7 9.94 5.15 2.54 8.47 4.16 1.95 0.93 0.46



Previous measurements from 3/24/05

1H db before cortab@probe after cortab@probe (W)with 13db attenuation
-6 7.7 8.5
-6 2.9 4.5
0 1.4 2.1
3 0.71 1.04
6 0.36 0.50

13C db @Amplifier 1 @probe (W) 8db attenution
-6 - 120.0
-3 - 59.0
0 - 29.0
3 - 15.0
6 - 7.3

15N db @Amplifier 1 @probe (W) 2db attenuation
-6 - 211 Linear till 6dB

2H db @Amplifier 1 @probe (W) 7db attenuation
-6 - 23 Linear till 6dB
  • Set ALLOWTOPICVIEW =

-- MichaelGoger - 13 Dec 2005

 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback