Difference: SerialEmParameters (1 vs. 36)

Revision 3629 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to dose
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)
Added:
>
>
  • Take an image, do ctrl i. The value for 'mean' is what you should write for target number of counts when doing Gain Reference
 

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =

-- MarienaSilvestry - 27 June 2011 | Modified from KdDerr r30 file

Revision 3527 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
Changed:
<
<
  • Add stage positions of areas to does
>
>
  • Add stage positions of areas to dose
 
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =

-- MarienaSilvestry - 27 June 2011 | Modified from KdDerr r30 file

Revision 3427 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =
Changed:
<
<
-- MarienaSilvestry - 27 June 2011 Modified from KdDerr
>
>
-- MarienaSilvestry - 27 June 2011 | Modified from KdDerr r30 file

Revision 3327 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =
Changed:
<
<
-- MarienaSilvestry - 27 June 2011
>
>
-- MarienaSilvestry - 27 June 2011 Modified from KdDerr

Revision 3227 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =
Changed:
<
<
-- Main.Mariena Silvestry - 27 June 2011
>
>
-- MarienaSilvestry - 27 June 2011

Revision 3127 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H
Added:
>
>

To start next Tilt Series

  1. Go back to 0 tilt (on tilt control menu)
  2. Go to next area (predose if necessary)
  3. If you've changed magnification, go back to desired mag
  4. Do autofocus
  5. Start new tilt series
 

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
  • Set ALLOWTOPICVIEW =
Changed:
<
<
-- KdDerr - 02 Oct 2009
>
>
-- Main.Mariena Silvestry - 27 June 2011

Revision 3024 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron
Changed:
<
<
  • When changing the binning value after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
>
>
  • When changing the binning value in Camera&Macro Ctrls after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
 
  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2924 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls and click Start (to start the montage)
  9. Save the montage as a (.st) stack file, double check the file extension (.st).
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron
Changed:
<
<
>
>
  • When changing the binning value after taking a full montage at 170x, don't forget to change the exposure time to 1 sec
 
  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2824 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
Changed:
<
<
  1. Go to Navigator Menu/Open, then ./Montage/Super Montage/Setup Full Montage, Click OK
>
>
  1. Go to Navigator Menu/Open, then Navigator/Montage/Super Montage/Setup Full Montage, Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  3. Go to Montage controls and click Start (to start the montage)
  4. Save the montage as a (.st) stack file, double check the file extension (.st).
  5. Save navigator file as .nav
  6. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  7. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2724 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
Changed:
<
<
  1. Go to Navigator Menu/Open, then Montage/Super Montage/Setup Full Montage, Click OK
>
>
  1. Go to Navigator Menu/Open, then ./Montage/Super Montage/Setup Full Montage, Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  3. Go to Montage controls and click Start (to start the montage)
  4. Save the montage as a (.st) stack file, double check the file extension (.st).
  5. Save navigator file as .nav
  6. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  7. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2624 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
Changed:
<
<
  1. Go to Navigator Menu/Open, then Menu/Montage/Super Montage/Setup Full Montage, Click OK
>
>
  1. Go to Navigator Menu/Open, then Montage/Super Montage/Setup Full Montage, Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  3. Go to Montage controls and click Start (to start the montage)
  4. Save the montage as a (.st) stack file, double check the file extension (.st).
  5. Save navigator file as .nav
  6. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  7. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2524 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Menu/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
Changed:
<
<
  1. Go to Montage controls windows and click Start (to start the montage)
>
>
  1. Go to Montage controls and click Start (to start the montage)
 
  1. Save the montage as a (.st) stack file, double check the file extension (.st).
  2. Save navigator file as .nav
  3. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  4. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2424 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X, maximum beam spread.
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator Menu/Open, then Menu/Montage/Super Montage/Setup Full Montage, Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
Changed:
<
<
  1. Go to Montage controls windows and click Start (to start the montage)
>
>
  1. Go to Montage controls windows and click Start (to start the montage)
 
  1. Save the montage as a (.st) stack file, double check the file extension (.st).
  2. Save navigator file as .nav
  3. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  4. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2324 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

Changed:
<
<
  1. Set magnification to 170X
>
>
  1. Set magnification to 170X, maximum beam spread.
 
  1. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  2. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  3. Make sure to increase registration number if you intend to transfer points from previous montages
  4. Go to Navigator Menu/Open, then Menu/Montage/Super Montage/Setup Full Montage, Click OK
  5. For the header of the montage file select all the available options
  6. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  7. Go to Montage controls windows and click Start (to start the montage)
  8. Save the montage as a (.st) stack file, double check the file extension (.st).
  9. Save navigator file as .nav
  10. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  11. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2224 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
Changed:
<
<
  1. Go to Navigator menu/Open/Montage/Super Montage/Setup Full Montage, Click OK
>
>
  1. Go to Navigator Menu/Open, then Menu/Montage/Super Montage/Setup Full Montage, Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
Changed:
<
<
  1. Go to Montage controls/Start (to start the montage)
  2. Save the montage as a (.st) stack file.
>
>
  1. Go to Montage controls windows and click Start (to start the montage)
  2. Save the montage as a (.st) stack file, double check the file extension (.st).
 
  1. Save navigator file as .nav
  2. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  3. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2124 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Contents

Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM


To make a montage of your grid

  1. Set magnification to 170X
Changed:
<
<
  1. Change spot size 5 (with the console buttons)
>
>
  1. Change spot size 5 (with the console buttons, check Tecnai interface on left screen)
 
  1. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  2. Make sure to increase registration number if you intend to transfer points from previous montages
Changed:
<
<
  1. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
>
>
  1. Go to Navigator menu/Open/Montage/Super Montage/Setup Full Montage, Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  3. Go to Montage controls/Start (to start the montage)
  4. Save the montage as a (.st) stack file.
  5. Save navigator file as .nav
  6. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  7. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 2024 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Added:
>
>
Contents
 Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1924 Nov 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
  12. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1803 Aug 2010 - Main.BillRice

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav
  11. When done, it will ask if you want to make a new map. ALWAYS SAY YES
Added:
>
>
  1. Choose file-->close, otherwise you will not be able to change camera settings. This only closes the file for writing.
 

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1703 Aug 2010 - Main.BillRice

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav
Added:
>
>
  1. When done, it will ask if you want to make a new map. ALWAYS SAY YES
 

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1606 Jul 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second
Added:
>
>

To Gain with Camera

  • Go to Camera/Gain Ref Policy(default4900)
 

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1528 Jun 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Predosing Sample

  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1428 Jun 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Changed:
<
<

Predosing Sample

>
>

Predosing Sample

 
  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1328 Jun 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Changed:
<
<

Predosing Sample

>
>

Predosing Sample

 
  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done
Deleted:
<
<
 

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1228 Jun 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 5 (with the console buttons)
  3. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  4. Make sure to increase registration number if you intend to transfer points from previous montages
  5. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  6. For the header of the montage file select all the available options
  7. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  8. Go to Montage controls/Start (to start the montage)
  9. Save the montage as a (.st) stack file.
  10. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Changed:
<
<
>
>

Predosing Sample

Added:
>
>
  • Go to 150 micron Condenser 2 aperture #4
  • Obtain a new Gain Reference to calibrate dose
    • Take record image use ctl + i to get stats
    • Camera/Prepare Gain Ref...
    • Enter mean value in target dose box
  • Add stage positions of areas to does
  • Check Acquire box for each point
  • Navigator/Acquire at points...
  • Select Macro 4 to dose 2000e-
  • Select Close Column Valves when done
 

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1128 Jun 2010 - Main.LilySun

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
Changed:
<
<
  1. Change spot size 4 (with the console buttons)
  2. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  3. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
>
>
  1. Change spot size 5 (with the console buttons)
  2. Go to Camera&Macro Controls Setup/Record tab Bining: 8 Exposure: 0.1 second
  3. Make sure to increase registration number if you intend to transfer points from previous montages
Added:
>
>
  1. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
 
  1. For the header of the montage file select all the available options
  2. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  3. Go to Montage controls/Start (to start the montage)
  4. Save the montage as a (.st) stack file.
  5. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 1009 Mar 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
  7. Go to Montage controls/Start (to start the montage)
  8. Save the montage as a (.st) stack file.
  9. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 903 Feb 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
Added:
>
>
  1. 10 % overlap approximately 11x11 pieces. Larger files will crash in loading.
 
  1. Go to Montage controls/Start (to start the montage)
  2. Save the montage as a (.st) stack file.
  3. Save navigator file as .nav

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations
Added:
>
>

Aligning rotated montage in Navigator

  1. Increase registration
  2. Collect new montage with identical parameters
  3. Select a minimum of 3 points on both maps
    • Magnifying the image when picking points is helpful to make the marked points more accurate.
  4. Go to each point and check Registration Point box on top of Navigator panel.
    • 1R1 = 2R1, 1R2 = 2R2, etc
  5. Go to Navigator on pull down menu and select Transform Items

 

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 803 Feb 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. Go to Montage controls/Start (to start the montage)
  7. Save the montage as a (.st) stack file.
  8. Save navigator file as .nav
Added:
>
>

Low Dose Montage

  1. Go to exposure mode in Tecnai Low Dose
  2. Open SerialEM
    • Once SerialEM is open do not use Tecnai interface to control microscope
  3. See above instructions
  4. After collecting montage pick points, move to area and save point in the Tecnai Specimen Position feature
  5. Close SerialEM
  6. Resume normal Low Dose operations

 

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 711 Jan 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. Go to Montage controls/Start (to start the montage)
  7. Save the montage as a (.st) stack file.
  8. Save navigator file as .nav

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Changed:
<
<
Magnification Binning Defocus (2009) Corrected Defocus (2010)
7800x 2 -8 micron -6.6 micron
9600x 2 -5.9 micron -4.2 micron
11500x 2 -4.1 micron -3.0 micron
14500x 2 -2.6 micron -1.8 micron
19000x 2 -2.3 micron -1.0 micron
25000x 2 -1.9 micron -0.67 micron
29000x 2 -0.9 micron -0.4 micron
>
>
Magnification Binning Defocus
7800x 2 -6.6 micron
9600x 2 -4.2 micron
11500x 2 -3.0 micron
14500x 2 -1.8 micron
19000x 2 -1.0 micron
25000x 2 -0.67 micron
29000x 2 -0.4 micron
 

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 611 Jan 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. Go to Montage controls/Start (to start the montage)
  7. Save the montage as a (.st) stack file.
  8. Save navigator file as .nav

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Changed:
<
<
Magnification Binning Defocus (2009 values slightly incorrect) Corrected Defocus (2010)
>
>
Magnification Binning Defocus (2009) Corrected Defocus (2010)
 
7800x 2 -8 micron -6.6 micron
9600x 2 -5.9 micron -4.2 micron
11500x 2 -4.1 micron -3.0 micron
14500x 2 -2.6 micron -1.8 micron
19000x 2 -2.3 micron -1.0 micron
25000x 2 -1.9 micron -0.67 micron
29000x 2 -0.9 micron -0.4 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 511 Jan 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. Go to Montage controls/Start (to start the montage)
  7. Save the montage as a (.st) stack file.
  8. Save navigator file as .nav

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus (2009 values slightly incorrect) Corrected Defocus (2010)
Changed:
<
<
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
>
>
7800x 2 -8 micron -6.6 micron
9600x 2 -5.9 micron -4.2 micron
11500x 2 -4.1 micron -3.0 micron
14500x 2 -2.6 micron -1.8 micron
19000x 2 -2.3 micron -1.0 micron
25000x 2 -1.9 micron -0.67 micron
29000x 2 -0.9 micron -0.4 micron
Deleted:
<
<
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron
 

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 411 Jan 2010 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
  3. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  4. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
  5. For the header of the montage file select all the available options
  6. Go to Montage controls/Start (to start the montage)
  7. Save the montage as a (.st) stack file.
  8. Save navigator file as .nav

Tomogram settings

  • File/close (to close montage image stack)
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second

To start a Tilt Series

  1. Go to Tilt Series menu/Setup/Start
  2. Select desired parameters

To finish a Tilt Series

  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice

F20 Defocus table for common magnifications

Changed:
<
<
Magnification Binning Defocus
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron
>
>
Magnification Binning Defocus (2009 values slightly incorrect) Corrected Defocus (2010)
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron
 

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 302 Oct 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Added:
>
>
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM
 

To make a montage of your grid

  1. Set magnification to 170X
  2. Change spot size 4 (with the console buttons)
Changed:
<
<
  1. Go to Setup menu  Record  Bin 8  exposure 0.1 second
  2. Go to Navigator menu  Open  Montage/Super Montage  Setup  full montage  Click OK
>
>
  1. Go to Setup menu/Record Bining: 8 Exposure: 0.1 second
  2. Go to Navigator menu/Open/Montage/Super Montage/Setupfull montage Click OK
 
  1. For the header of the montage file select all the available options
Changed:
<
<
  1. Go to Montage controls  Start (to start the montage)
  2. Save the montage as a stack file.
>
>
  1. Go to Montage controls/Start (to start the montage)
  2. Save the montage as a (.st) stack file.
Added:
>
>
  1. Save navigator file as .nav
 
Changed:
<
<

For tomography

  • select a bining of 2 and a 1 second exposure
>
>

Tomogram settings

  • File/close (to close montage image stack)
Added:
>
>
  • Go to Setup menu/Record Bining: 2 Exposure: 1 second
 

To start a Tilt Series

Changed:
<
<
>
>
  1. Go to Tilt Series menu/Setup/Start
Deleted:
<
<
  1. Go to Tilt Series menu  Setup/Start
 
  1. Select desired parameters

To finish a Tilt Series

Added:
>
>
  1. After the tilt series is done go to Tilt Series menu/Terminate
  2. Select "Yes" to close image stack
  3. Select "No" to save 90 degrees rotated image in Buffer H
 
Deleted:
<
<
  1. After the tilt series is done go to Tilt Series menu  Terminate
  2. Select Yes in the message on your screen
  3. Select No to save 90 degrees rotated
 
Deleted:
<
<
 

Useful Tip

Changed:
<
<
  • If you want to go to a particular slice of your stack file go to the File menu  Read  select file  select slice
>
>
  • If you want to go to a particular slice of your stack file go to the File menu/Read/select file/select slice
 

F20 Defocus table for common magnifications

Magnification Binning Defocus
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 202 Oct 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Added:
>
>

To make a montage of your grid

 
Changed:
<
<

When you start with the TF20

>
>
  1. Set magnification to 170X
Added:
>
>
  1. Change spot size 4 (with the console buttons)
  2. Go to Setup menu  Record  Bin 8  exposure 0.1 second
  3. Go to Navigator menu  Open  Montage/Super Montage  Setup  full montage  Click OK
  4. For the header of the montage file select all the available options
  5. Go to Montage controls  Start (to start the montage)
  6. Save the montage as a stack file.
 
Deleted:
<
<
1- Check that the IGP1 value is below 15 then is ready to use 2- Drop the screen with some of the console buttons (check Serial EM for the function of each of the buttons)
 
Changed:
<
<
To make a montage of your grid
>
>

For tomography

Added:
>
>
  • select a bining of 2 and a 1 second exposure

To start a Tilt Series

 
Changed:
<
<
1- Set magnification to 170X 2- Change spot size 4 (with the console buttons)
>
>
  1. Go to Tilt Series menu  Setup/Start
  2. Select desired parameters
Deleted:
<
<
3- Go to Setup menu  Record  Bin 8  exposure 0.1 second 4- Go to Navigator menu  Open  Montage/Super Montage  Setup  full montage  Click OK 5- For the header of the montage file select all the available options 6- Go to Montage controls  Start (to start the montage) 7- Save the montage as a stack file.
 
Added:
>
>

To finish a Tilt Series

 
Changed:
<
<
For tomography select a bining of 2 and a 1 second exposure
>
>
  1. After the tilt series is done go to Tilt Series menu  Terminate
Added:
>
>
  1. Select Yes in the message on your screen
  2. Select No to save 90 degrees rotated
 
Changed:
<
<
To start a Tilt Series
>
>

Useful Tip

  • If you want to go to a particular slice of your stack file go to the File menu  Read  select file  select slice
Deleted:
<
<
1- Go to Tilt Series menu  Setup/Start 2- Select desired parameters

To finish a Tilt Series

1- After the tilt series is done go to Tilt Series menu  Terminate 2- Select Yes in the message on your screen 3- Select No to save 90 degrees rotated

Useful Tip

If you want to go to a particular slice of your stack file go to the File menu  Read  select file  select slice

 

F20 Defocus table for common magnifications

Magnification Binning Defocus
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

Revision 102 Oct 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"

When you start with the TF20

1- Check that the IGP1 value is below 15 then is ready to use 2- Drop the screen with some of the console buttons (check Serial EM for the function of each of the buttons)

To make a montage of your grid

1- Set magnification to 170X 2- Change spot size 4 (with the console buttons) 3- Go to Setup menu  Record  Bin 8  exposure 0.1 second 4- Go to Navigator menu  Open  Montage/Super Montage  Setup  full montage  Click OK 5- For the header of the montage file select all the available options 6- Go to Montage controls  Start (to start the montage) 7- Save the montage as a stack file.

For tomography select a bining of 2 and a 1 second exposure

To start a Tilt Series

1- Go to Tilt Series menu  Setup/Start 2- Select desired parameters

To finish a Tilt Series

1- After the tilt series is done go to Tilt Series menu  Terminate 2- Select Yes in the message on your screen 3- Select No to save 90 degrees rotated

Useful Tip

If you want to go to a particular slice of your stack file go to the File menu  Read  select file  select slice

F20 Defocus table for common magnifications

Magnification Binning Defocus
3500x 2 -43 micron
3500x 1 -11.2 micron
7800x 2 -8 micron
7800x 1 -2.3 micron
9600x 2 -5.9 micron
9600x 1 -1.4 micron
11500x 2 -4.1 micron
11500x 1 -1 micron
14500x 2 -2.6 micron
14500x 1 -0.6 micron
15000x 2 -2.3 micron
15000x 1 -0.6 micron
17000x 2 -1.9 micron
25000x 2 -0.9 micron

  • Set ALLOWTOPICVIEW =

-- KdDerr - 02 Oct 2009

 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback