Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Register use with pHp
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open vacuum overview (bottom right)
Changed:
<
<
    • IGPI should not read over 15 (6 best value)
>
>
    • IGP1 should not read over 15 (6 best value)
 
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2810 Jun 2011 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

Changed:
<
<
  • Log on to TEM data computer (left screen)
>
>
  • Register use with pHp
Deleted:
<
<
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
 
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
Deleted:
<
<
  • Open Tecnai User Interface
 
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2710 Dec 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Tecnai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<
>
>
 

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2610 Dec 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Tecnai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<
>
>
 

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2510 Dec 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Tecnai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<
>
>
 

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2410 Dec 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
Changed:
<
<
  • Open Technai User Interface
>
>
  • Open Tecnai User Interface
 
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<
>
>
 

Check for beam and alignment before loading sample (NYSBC Staff)

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2324 Nov 2009 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<

Check for beam and alignment before loading sample

>
>

Check for beam and alignment before loading sample (NYSBC Staff)

 
  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2231 Dec 2008 - Main.BillRice

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Changed:
<
<
>
>

Revision 2123 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors

Common Problems

Instructions to Locate Beam

Added:
>
>
 

Instructions to Find CCD Blank Enable

Changed:
<
<

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
>
>
Added:
>
>

Acessing status of the FEG

  • Open Tecnai User Interface
  • Go to Setup/FEG Control panel
    • If the operate button is not yellow FEG is off
    • Call NYSBC or FEI staff

Keeping the beam centered when raising screen (as opposed to offsetting for the camera):

 
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 2023 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

Changed:
<
<
    • Insert specimen holder to first stop
    • In setup menu open column valves when option becomes available
    • Start Low Dose mode before aligning the beam or alignments will be off.
>
>
  • Insert specimen holder to first stop
  • In setup menu open column valves when option becomes available
  • Start Low Dose mode before aligning the beam or alignments will be off.
 
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
Added:
>
>
  • Warm up cryoholder if used
  • Log out of computers and turn off monitors
 

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1923 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

Changed:
<
<
TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software
>
>
  • TEM data computer controls Fast Scan CCD holds data and provides access to internet
  • FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software
 Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves when option becomes available
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1822 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves when option becomes available
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button

Sample Insertion

  • Have NYSBC staff insert your sample

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Changed:
<
<

Objective Aperture Alignment and Stigmation

>
>

Objective Aperture Alignment and Stigmation

 
  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button
Changed:
<
<

Adjust Eucentric Focus

>
>

Adjust Eucentric Focus

 
  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”
Changed:
<
<

Computer Crash

>
>
Deleted:
<
<
  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
    • right click and select manage/services and applications/services/TEM server
    • Press start
  • Logon as yourself
 

Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1722 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam and alignment before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves when option becomes available
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Check Condenser Aperture Astigmatism
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
    • Use multifunction knob to make beam round
    • Unclick Stigmator button
Changed:
<
<

InsertRoomTemperatureSample

InsertCryoSample

>
>

Sample Insertion

  • Have NYSBC staff insert your sample
 

Specimen Imaging

  • Check vacuum IGP1 should be max 15
  • Open column valves

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
  • Move Z +/- to obtain focus
    • if you are having trouble seeing thon rings dropping the magnification may be helpful

  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Reset holder
  • Remove objective aperture
  • Remove sample
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”

Computer Crash

  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
    • right click and select manage/services and applications/services/TEM server
    • Press start
  • Logon as yourself


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1622 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
Deleted:
<
<
    • If necessary, CreateNewUser?
 
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<

Check for beam and set up before loading sample

>
>
Added:
>
>

Check for beam and alignment before loading sample

 
    • Insert specimen holder to first stop
Changed:
<
<
    • In setup menu open column valves
>
>
    • In setup menu open column valves when option becomes available
 
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
Changed:
<
<
* Condenser Aperture Alignment and Stigmation
>
>
  • Condenser Aperture Alignment and Stigmation
 
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
Changed:
<
<
  • Press Stigmator button on left control panel
>
>
  • Check Condenser Aperture Astigmatism
Added:
>
>
    • Bring beam to crossover; if beam is not round
    • Press Stigmator button on left control panel
 
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
Changed:
<
<
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button
>
>
    • Use multifunction knob to make beam round
    • Unclick Stigmator button
 

InsertRoomTemperatureSample

InsertCryoSample

Changed:
<
<

Setup Beam

>
>

Specimen Imaging

 
  • Check vacuum IGP1 should be max 15
  • Open column valves
Deleted:
<
<
  • Obtain rough focus
 

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button
Changed:
<
<

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
>
>

Adjust Eucentric Focus

  • Press Eucentric Focus on Right control panel
Deleted:
<
<
  • Click Alpha Wobbler L2
 
  • Move Z +/- to obtain focus
Changed:
<
<

A few words about focus

>
>
    • if you are having trouble seeing thon rings dropping the magnification may be helpful
Deleted:
<
<
  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode
 
Added:
>
>
  • Alpha Wobbler method
    • in Search mode
    • press alpha wobbler Left control panel
    • Move Z +/- to obtain focus
 

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
Changed:
<
<
  • Turn on turbo
>
>
  • Reset holder
 
  • Remove objective aperture
Added:
>
>
  • Remove sample
 
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”

Computer Crash

  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
    • right click and select manage/services and applications/services/TEM server
    • Press start
  • Logon as yourself


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1522 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
Changed:
<
<
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
>
>
  • Register film number with pHp
  • Log on to TEM control computer (right two screens)
 
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked
Changed:
<
<

Check for beam before loading sample

>
>

Check for beam and set up before loading sample

 
    • Insert specimen holder to first stop
    • In setup menu open column valves
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
Changed:
<
<

InsertRoomTemperatureSample

>
>
* Condenser Aperture Alignment and Stigmation
Deleted:
<
<

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
 
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button
Added:
>
>

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGP1 should be max 15
  • Open column valves
  • Obtain rough focus

 

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”

Computer Crash

  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
    • right click and select manage/services and applications/services/TEM server
    • Press start
  • Logon as yourself


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1417 Jul 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”

Computer Crash

  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
Changed:
<
<
    • right click and select manage\services and applications\services\TEM server
>
>
    • right click and select manage/services and applications/services/TEM server
 
    • Press start
  • Logon as yourself


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1317 Jul 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • Start Low Dose mode before aligning the beam or alignments will be off.
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”
Added:
>
>

Computer Crash

  • Close gun valve manually
    • This is located on the left side of the viewing port labeled "Close V7"
    • If you manually close the valves you must reopen them before clicking "Open Column Valves"
  • Restart computer
  • Logon as Administrator
  • Goto: My Computer
    • right click and select manage\services and applications\services\TEM server
    • Press start
  • Logon as yourself
 

Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1216 Jul 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • Start Low Dose mode before aligning the beam or alignments will be off.
Changed:
<
<
    • If no beam lower magnification
>
>
    • If no beam go to Focus and lower magnification, do not adjust the beam in Search
 

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1116 Jul 2007 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
Added:
>
>
    • Start Low Dose mode before aligning the beam or alignments will be off.
 
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in em-signup database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 1014 Nov 2006 - Main.BillRice

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
Changed:
<
<
  • Record date and ending plate number in Grey Matter database on TEM data computer
>
>
  • Record date and ending plate number in em-signup database on TEM data computer
 

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 904 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Changed:
<
<

FindBeam

CcdBlankEnable

>
>

Instructions to Locate Beam

Instructions to Find CCD Blank Enable

 

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 804 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get close to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)

Common Problems

Added:
>
>

FindBeam

CcdBlankEnable

 

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 704 Oct 2006 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
  • CCD Blank enable should be on screen with box checked

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

Changed:
<
<
  • Get clost to focus with focus knob (right panel)
>
>
  • Get close to focus with focus knob (right panel)
 
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)
Added:
>
>

Common Problems

Keeping the beam centered when raising screen

(as opposed to offsetting for the camera):
  • Workspace Layout
    • Detector Config
    • Uncheck “Auto Mode”
 

Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 613 Sep 2006 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder
Changed:
<
<
*CCD Blank enable should be on screen with box checked Instructions to Retreive *CCDBlankEnable Box
>
>
 

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get clost to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 513 Sep 2006 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Changed:
<
<
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)
    • While airlock is prepumping load sample in holder

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get clost to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

>
>
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read ~3900-3950
  • Accelerating voltage should be 200kV
  • Click Prepump airlock (setup menu)

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Get clost to focus with focus knob (right panel)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>

Revision 401 Aug 2005 - Main.KdDerr

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
Changed:
<
<
  • Extraction voltage should read 3900
>
>
  • Extraction voltage should read ~3900-3950
 
  • Accelerating voltage should be 200kV
Changed:
<
<
  • Activate turbo pump on airlock (setup menu)
>
>
  • Click Prepump airlock (setup menu)
 
    • While airlock is prepumping load sample in holder

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

Deleted:
<
<
  • Turn off airlock turbo pump
 
  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

Changed:
<
<
  • Click Eucentric focus (right control pad)
>
>
  • Get clost to focus with focus knob (right panel)
Deleted:
<
<
  • Find focus with focus knob (right control pad)
 
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 322 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
    • If necessary, CreateNewUser?
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read 3900
  • Accelerating voltage should be 200kV
  • Activate turbo pump on airlock (setup menu)
    • While airlock is prepumping load sample in holder

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Turn off airlock turbo pump
  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Click Eucentric focus (right control pad)
  • Find focus with focus knob (right control pad)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>

Revision 218 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
Added:
>
>
    • If necessary, CreateNewUser?
 
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read 3900
  • Accelerating voltage should be 200kV
  • Activate turbo pump on airlock (setup menu)
    • While airlock is prepumping load sample in holder

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Turn off airlock turbo pump
  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Click Eucentric focus (right control pad)
  • Find focus with focus knob (right control pad)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode
Deleted:
<
<

TomoGraphy

 

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 118 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="CryoEM"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tecnai F20 (basic operations)

TEM data computer controls Fast Scan CCD holds data and provides access to internet FEI control computer controls EM and Slow Scan 4k CCD camera and runs tomography software

Press Scroll Lock 2x to switch screens

Contents

Initial Setup

  • Log on to TEM data computer (left screen)
  • Register use with Grey matter
  • Log on to TEM control computer (right double screen)
  • Start TCL
    • Click start in dialog box
    • EMMENU (pull down)
    • Select Start EMMenu
  • Open Technai User Interface
  • Open vacuum overview (bottom right)
    • IGPI should not read over 15 (6 best value)
    • P3 indicates moisture of film (60 new film-20 dry film)
  • Extraction voltage should read 3900
  • Accelerating voltage should be 200kV
  • Activate turbo pump on airlock (setup menu)
    • While airlock is prepumping load sample in holder

Check for beam before loading sample

    • Insert specimen holder to first stop
    • In setup menu open column valves
    • If no beam lower magnification

InsertRoomTemperatureSample

InsertCryoSample

Setup Beam

  • Turn off airlock turbo pump
  • Check vacuum IGPI should be max 15
  • Open column valves
  • Obtain rough focus
  • Condenser Aperture Alignment and Stigmation
  • Check condenser aperture size (plastic sections-#4)
  • Check alignment of condenser aperture
    • At cross over center beam with beam shift (left control pad)
    • Spread beam and center with aperture X, Y
    • Repeat until beam spreads evenly
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Condenser button is selected
  • Bring beam to crossover
  • Use multifunction knob to make beam round
  • Unclick Stigmator button

Objective Aperture Alignment and Stigmation

  • Check Objective aperture size (plastic sections-#3)
  • Put in objective aperture
    • Press diffraction (right control pad)
    • Condense diffraction spot
  • Center objective aperture around the spot with the aperture controls X, Y
  • Press Stigmator button on left control panel
    • Stigmator menu should pop up
    • Make sure Objective button is selected
  • Using CCD camera and Power spectrum bring image to focus
    • If Thon rings are out of round fix with multifunction knobs
  • Unclick Stigmator button

Adjust Eucentric Specimen Height

  • Click Eucentric focus (right control pad)
  • Find focus with focus knob (right control pad)
  • Go to low magnification
  • Click Alpha Wobbler L2
  • Move Z +/- to obtain focus

A few words about focus

  • If you are using the power spectra (easier to use at higher magnifications)
  • Over-focus has a bright center with division ring
  • Focus has a bright center that lessens in intensity as it moves from center
  • Under-focus has a bright center with division rings forming from the outside
  • The amount of under focus used for an image is directly dependant on the size of the feature of interest.

Low Dose

  • Click low dose tab
  • Use search to find area of interest
  • Use focus to focus area of interest
  • When ready to image use exposure mode

TomoGraphy

Taking a FilmImage

Taking a CcdImage

Shut down

  • Close column valves
  • Turn on turbo
  • Remove objective aperture
  • Record date and ending plate number in Grey Matter database on TEM data computer

End of day shutdown

  • Remove Dewar for cold trap and place the cup under the coils
  • Run cryo cycle (setup menu)


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM