Difference: TietzTomography (1 vs. 8)

Revision 810 Aug 2009 - Main.BillRice

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Changed:
<
<

This Software Is No Longer In Use at NYSBC

>
>

This Software Is No Longer In Use at NYSBC : Use SerialEM

 

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations

Update Flat Field images (if necessary)

  • Remove sample to first stop or Trash.findDFdf a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable

Calibrate focus (always necessary)

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Calibrations (if necessary)

  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used

Check Tilt Range

  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Open series in EM Menu
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 710 Aug 2009 - Main.BillRice

 
META TOPICPARENT name="TomoGraphy"
Changed:
<
<
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations

Update Flat Field images (if necessary)

  • Remove sample to first stop or Trash.findDFdf a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable

Calibrate focus (always necessary)

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Calibrations (if necessary)

  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used

Check Tilt Range

  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Open series in EM Menu
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

>
>
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

This Software Is No Longer In Use at NYSBC

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations

Update Flat Field images (if necessary)

  • Remove sample to first stop or Trash.findDFdf a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable

Calibrate focus (always necessary)

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Calibrations (if necessary)

  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used

Check Tilt Range

  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Open series in EM Menu
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>

Revision 617 Sep 2008 - Main.DavidCowburn

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations

Update Flat Field images (if necessary)

  • Remove sample to first stop or Trash.findDFdf a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable

Calibrate focus (always necessary)

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Calibrations (if necessary)

  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used

Check Tilt Range

  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Open series in EM Menu
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 516 Jul 2008 - Main.DavidCowburn

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Check Eucentric Focus

Changed:
<
<
  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus
>
>
  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus
 

Microscope alignment

Changed:
<
<
  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations
>
>
  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
  • Skip cosmetic alignments and calibrations
 

Update Flat Field images (if necessary)

Changed:
<
<
  • Remove sample to first stop or find a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable
>
>
  • Remove sample to first stop or Trash.findDFdf a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable
 

Calibrate focus (always necessary)

Changed:
<
<
  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus
>
>
  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus
 

Start Tomography program

Changed:
<
<
  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography
>
>
  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography
 

Update Calibrations (if necessary)

Changed:
<
<
  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used
>
>
  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used
 

Check Tilt Range

Changed:
<
<
  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface
>
>
  • Locate area of interest
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface
 

Set up Tomography

Changed:
<
<
  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail
>
>
  • Refer to extensive Help section in EM Menu about tomography
  • In Positions dialog box
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail
 

Low Dose Tomography Setup

Changed:
<
<
  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive
>
>
  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive
 

Tilt Series

Changed:
<
<
  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start
>
>
  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start
 

Displaying series

Changed:
<
<
  • Click Open series in EM Menu
  • Use Series button to display as a movie
>
>
  • Click Open series in EM Menu
  • Use Series button to display as a movie
 


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Changed:
<
<
>
>
 

Revision 416 Jun 2005 - Main.KdDerr

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

Changed:
<
<
  • Essential to functioning of Autofocus and tomography in general
>
>
  • Essential to functioning of Autofocus and tomography in general found in Alignments workspace under Tune tab
 
  • Gun alignments (tilt and shift)
Changed:
<
<
  • Beam alignments (rotation center, pivot points, coma free)
  • Img alignments (pivot points)
>
>
  • Beam alignments (rotation center, pivot points, HM Beam)
  • Img alignments (HM Beam)
 
  • Skip cosmetic alignments and calibrations
Changed:
<
<

Start Tomography program

>
>

Update Flat Field images (if necessary)

Deleted:
<
<
  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Flat Field images if necessary

 
  • Remove sample to first stop or find a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable
Changed:
<
<
>
>

Calibrate focus (always necessary)

Deleted:
<
<

Calibrate focus

 
  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus
Changed:
<
<

Calibrate positions (if necessary)

  • Click Positions button
  • Click Calibrations
>
>

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
Added:
>
>
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Calibrations (if necessary)

  • Click Calibrations
  • Check dates of calibrations at all mags to be used
    • Click Update TEM after changing mag to get dates
 
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used
Deleted:
<
<
 

Check Tilt Range

Added:
>
>
  • Locate area of interest
 
  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
Changed:
<
<
  • Click Positions Button
>
>
  • In Positions dialog box
 
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

Changed:
<
<
  • Click Load series in EM Menu
>
>
  • Click Open series in EM Menu
 
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 322 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Contents

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, coma free)
  • Img alignments (pivot points)
  • Skip cosmetic alignments and calibrations

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Update Flat Field images if necessary

  • Remove sample to first stop or find a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable

Calibrate focus

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus

Calibrate positions (if necessary)

  • Click Positions button
  • Click Calibrations
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used

Check Tilt Range

  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • Click Positions Button
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Load series in EM Menu
  • Use Series button to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>

Revision 218 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Changed:
<
<
Tomography using TVIPS EM Menu 3.0
>
>
Tomography using TVIPS EM Menu 3.0
 
Added:
>
>
Contents
 

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, coma free)
  • Img alignments (pivot points)
  • Skip cosmetic alignments and calibrations

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography
Changed:
<
<

Consider Flat Field images are already stored.

>
>

Update Flat Field images if necessary

Added:
>
>
  • Remove sample to first stop or find a hole
  • Use "Continuous Readout" to set illumination
    • End with spacebar
  • Provide stong beam for ~4000 counts average
  • Click Create Flat fields
  • Check data to make sure it is reasonable
 
Added:
>
>
 

Calibrate focus

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
Changed:
<
<
    • If blurred image is obtained during autofocus
      • could be image movement due to EM slow response
      • Go closer to focus
      • Weaken beam and lengthen exposure
      • Check focus
>
>
  • If blurred image is obtained during autofocus
    • could be image movement due to EM slow response
    • Go closer to focus
    • Weaken beam and lengthen exposure
    • Check focus
 
Added:
>
>

Calibrate positions (if necessary)

  • Click Positions button
  • Click Calibrations
  • ensure high contrast area with fairly bright beam and stable specimen
    • Top to bottom, click buttons
    • Calibrate for all mags to be used
 

Check Tilt Range

  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • Click Positions Button
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
Changed:
<
<
Check each position with exposure at maximum angles before beginning series Check "Lock Position" box Before center detail
>
>
  • Check each position with exposure at maximum angles before beginning series
  • Check Lock Position box
  • Before collecting series, click center detail
Deleted:
<
<
Before beginning series
 
Changed:
<
<
Tilt Series If start angle is positive Angle increments need to be negative If using "Saxton Screen" Start angle needs to be positive Check tracking box for all images Do not check Focus box for all images (about every third) Do not check search box for any images Press "Start"
>
>

Low Dose Tomography Setup

  • Go into Scanning position
  • Click Areas button
    • Allows definition of location of Tracking and Focus positions along the tilt axis
  • calculates offset of image shift
  • requires manual adjustment of beam shift
    • uncheck Lock Positions
    • Goto each position (Exposure, Tracking, Focus, Search) and adjust beam position
    • click Update Positions
Added:
>
>
  • Do not run Areas again
    • The image shift dislocations become additive, but not shown as additive
 
Changed:
<
<
Displaying series "Load series" in EMMenu Use "series" to display as a movie
>
>

Tilt Series

  • If start angle is positive angle increments need to be negative
  • If using Saxton Screen, start angle needs to be positive
Added:
>
>
  • Check tracking box for all images
  • Do not check Focus box for all images (about every third)
  • Do not check search box for any images
  • Press Start

Displaying series

  • Click Load series in EM Menu
  • Use Series button to display as a movie
 


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 118 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="TomoGraphy"
Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM


Tomography using TVIPS EM Menu 3.0

Check Eucentric Focus

  • Click "Eucentric focus" (right control pad)
  • Find focus with "focus" knob (right control pad)
  • Go to low magnification
  • Click "Alpha Wobbler" L2
  • Move Z +/- to obtain focus

Microscope alignment

  • Essential to functioning of Autofocus and tomography in general
  • Gun alignments (tilt and shift)
  • Beam alignments (rotation center, pivot points, coma free)
  • Img alignments (pivot points)
  • Skip cosmetic alignments and calibrations

Start Tomography program

  • Start TCL
  • Select EMMENU in pulldown
  • Click Positions
  • Read extensive Help section in EM Menu about tomography

Consider Flat Field images are already stored.

Calibrate focus

  • Set up high contrast image with fairly bright beam (gold particles good) close to focus
  • Click Auto focus
  • Adjust camera parameters for strong exposure 2x2 binning (e.g. 4000 counts)
  • click Calibrate
  • check calibration by selecting particular focus on upper left
    • check power spectrum and fitctf rings if >29k
  • Run for all magnifications you will be using
    • exposure and focus positions should be at the same mag so no focus differential
    • also no mag changes between focus and exposure (e.g., if tracking is done in between)
    • If blurred image is obtained during autofocus
      • could be image movement due to EM slow response
      • Go closer to focus
      • Weaken beam and lengthen exposure
      • Check focus

Check Tilt Range

  • Open Control TEM on EM Menu
  • Check either tilt extreme
  • Can also use TEM control feature in Technai User Interface

Set up Tomography

  • Refer to extensive Help section in EM Menu about tomography
  • Click Positions Button
  • Calibrations are already stored and do not need to be repeated camera has been moved or goiniometer has been worked on
  • Click Show Positions button to check parameters for each of the 4 "Positions"
    • Set camera parameters for each position (format and exposure)
    • Set defocus under Focus tab
  • Click apply after any change
  • Go to lower magnification (~ 4 clicks)
  • Update Scanning parameters and copy to Search
  • Go to Exposure Tracking Focus and adjust parameters
    • Press Update Positions when done
Check each position with exposure at maximum angles before beginning series Check "Lock Position" box Before center detail Before beginning series

Tilt Series If start angle is positive Angle increments need to be negative If using "Saxton Screen" Start angle needs to be positive Check tracking box for all images Do not check Focus box for all images (about every third) Do not check search box for any images Press "Start"

Displaying series "Load series" in EMMenu Use "series" to display as a movie


Visit the Cryo EM Website at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback