Difference: WohlwendUse (5 vs. 6)

Revision 605 Feb 2013 - Main.MarienaSilvestry

 
META TOPICPARENT name="HighPressFreezer"
Return to Cryo EM web page at http://www.nysbc.org/facilities/CEM

Principles & Protocols



Protocols - Wohlwend

Jump to Principles

Start up

  • Open nitrogen gas valve
  • Press yellow "System Dry Up" button. Run gas through for 30 minutes.
Added:
>
>
  • Turn off "System Dry Up."
 
  • Press green "On/Off" button
  • Press green "Nitrogen" button
  • Open liquid nitrogen tank
  • Wait until red "Nitrogen Level" light goes out
  • Press yellow "Drive in" button.
    • Yellow light will start flashing
    • "Piston Down" red lamp will go on
    • Oil pressure will increase to 100 bar
    • "Drive In" light will go out
  • Press green "Auto" button.

Use

  • Insert holder and safety pin
  • Green "Jet Ready" lamp should light
  • Press yellow "Jet Auto" button
  • Wait for yellow "Jet Over" lamp
  • Pull pin and remove sample

Measuring temperature and pressure (traces to check performance)

See attached document:
Changed:
<
<
* How_to_record_temperature_and_pressure_traces_on_the_Wohlwend_HPF.docx: SOP on recording temperature and maintenance on the WW HPF
>
>
 

Shut down

  • Close liquid nitrogen tank
  • Press green "On/Off" Button
  • Press yellow "System Dry Up" button * Light will blink for 10 hours while system drys
  • Once lamp stops flashing close nitrogen gas valve

  • Set ALLOWTOPICVIEW =

-- KdDerr - 03 Oct 2007

Changed:
<
<
META FILEATTACHMENT attachment="How_to_record_temperature_and_pressure_traces_on_the_Wohlwend_HPF.docx" attr="" comment="SOP on recording temperature and maintenance on the WW HPF" date="1358544928" name="How_to_record_temperature_and_pressure_traces_on_the_Wohlwend_HPF.docx" path="How to record temperature and pressure traces on the Wohlwend HPF.docx" size="321108" stream="How to record temperature and pressure traces on the Wohlwend HPF.docx" user="Main.MarienaSilvestry" version="1"
>
>
META FILEATTACHMENT attachment="How_to_record_temperature_and_pressure_traces_on_the_Wohlwend_HPF.docx" attr="" comment="Recording press and temp maintenance on WW HPF" date="1360104102" name="How_to_record_temperature_and_pressure_traces_on_the_Wohlwend_HPF.docx" path="How to record temperature and pressure traces on the Wohlwend HPF.docx" size="321475" stream="How to record temperature and pressure traces on the Wohlwend HPF.docx" user="Main.MarienaSilvestry" version="1"
 
 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback