Difference: CryoUltraMicrotomy (1 vs. 8)

Revision 801 Sep 2009 - Main.KdDerr

Changed:
<
<
META TOPICPARENT name="SamplePreparation"
>
>
META TOPICPARENT name="CryoSP"
 CryoUltraMicrotomy


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 710 Aug 2009 - Main.BillRice

 
META TOPICPARENT name="SamplePreparation"
Changed:
<
<
CryoUltraMicrotomy

*Microtome Conversion

*Cryosectioning notes from Chyongere Hirsch (Wadsworth Center Albany,NY)24 Mar 05


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

>
>
CryoUltraMicrotomy


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Deleted:
<
<

Revision 608 Jun 2005 - Main.KdDerr

 
META TOPICPARENT name="SamplePreparation"
Changed:
<
<
Microtome Conversion
>
>
CryoUltraMicrotomy
 
Changed:
<
<

>
>
*Microtome Conversion
Deleted:
<
<
Contents
 
Changed:
<
<

Remove Cryosectioning unit

>
>
*Cryosectioning notes from Chyongere Hirsch (Wadsworth Center Albany,NY)24 Mar 05
Deleted:
<
<
  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself

Attaching room temperature sectioning unit

  • Slide plastic sectioning tray until plastic is flush with microtome frame

Removing room temperature unit

  • slide plastic tray towards yourself

Attaching cryosectioning unit

  • Raise specimen arm to top position
  • Slide Cryo unit onto tracks
  • Connect sectioning arm with specimen holder
  • Slide cryo unit remainder of the way on
  • Fasten specimen arm thumb screw
  • Make sure hand wheel can make a complete turn
  • Fasten two levers on right
  • Give a light tug to make sure they are secure
  • Fasten cable on left to back of machine
  • Connect LN2 hose
  • Insert specimen chuck
  • Insert knife holder
  • Turn on grey control unit
  • Cool down
 

Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 522 Apr 2005 - Main.KdDerr

 
META TOPICPARENT name="SamplePreparation"
Changed:
<
<
Cryo Ultramicrotomy
>
>
Microtome Conversion
 
Contents

Remove Cryosectioning unit

  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself

Attaching room temperature sectioning unit

  • Slide plastic sectioning tray until plastic is flush with microtome frame

Removing room temperature unit

  • slide plastic tray towards yourself

Attaching cryosectioning unit

  • Raise specimen arm to top position
  • Slide Cryo unit onto tracks
  • Connect sectioning arm with specimen holder
  • Slide cryo unit remainder of the way on
  • Fasten specimen arm thumb screw
  • Make sure hand wheel can make a complete turn
  • Fasten two levers on right
  • Give a light tug to make sure they are secure
  • Fasten cable on left to back of machine
  • Connect LN2 hose
  • Insert specimen chuck
  • Insert knife holder
  • Turn on grey control unit
  • Cool down


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 422 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="SamplePreparation"
Cryo Ultramicrotomy


Contents

Remove Cryosectioning unit

  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself

Attaching room temperature sectioning unit

  • Slide plastic sectioning tray until plastic is flush with microtome frame

Removing room temperature unit

  • slide plastic tray towards yourself

Attaching cryosectioning unit

  • Raise specimen arm to top position
  • Slide Cryo unit onto tracks
  • Connect sectioning arm with specimen holder
  • Slide cryo unit remainder of the way on
  • Fasten specimen arm thumb screw
  • Make sure hand wheel can make a complete turn
  • Fasten two levers on right
  • Give a light tug to make sure they are secure
  • Fasten cable on left to back of machine
  • Connect LN2 hose
  • Insert specimen chuck
  • Insert knife holder
  • Turn on grey control unit
  • Cool down


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>

Revision 318 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="SamplePreparation"
Cryo Ultramicrotomy
Changed:
<
<

Cryo to room temperature sectioning conversion

>
>

Added:
>
>
Contents

Remove Cryosectioning unit

 
  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself
Changed:
<
<

Attaching room temperature sectioning unit

>
>

Attaching room temperature sectioning unit

 
  • Slide plastic sectioning tray until plastic is flush with microtome frame
Added:
>
>

Removing room temperature unit

  • slide plastic tray towards yourself

Attaching cryosectioning unit

  • Raise specimen arm to top position
  • Slide Cryo unit onto tracks
  • Connect sectioning arm with specimen holder
  • Slide cryo unit remainder of the way on
  • Fasten specimen arm thumb screw
  • Make sure hand wheel can make a complete turn
  • Fasten two levers on right
  • Give a light tug to make sure they are secure
  • Fasten cable on left to back of machine
  • Connect LN2 hose
  • Insert specimen chuck
  • Insert knife holder
  • Turn on grey control unit
  • Cool down
 


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 218 Mar 2005 - Main.DavidStokes

Changed:
<
<
META TOPICPARENT name="CemfacGroup"
>
>
META TOPICPARENT name="SamplePreparation"
 Cryo Ultramicrotomy

Cryo to room temperature sectioning conversion

  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself

Attaching room temperature sectioning unit

  • Slide plastic sectioning tray until plastic is flush with microtome frame


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

Changed:
<
<
-- DavidStokes - 18 Mar 2005
>
>
-- DavidStokes - 18 Mar 2005
Deleted:
<
<

Revision 118 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="CemfacGroup"
Cryo Ultramicrotomy

Cryo to room temperature sectioning conversion

  • Bake for at least one hour before removing
  • Loosen thumb screw on specimen arm
  • Loosen both levers on right side
  • Unplug cable of left
  • Slide cryo unit towards yourself

Attaching room temperature sectioning unit

  • Slide plastic sectioning tray until plastic is flush with microtome frame


Visit the Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback