Difference:
GlowDischarge
(11 vs. 12)
Revision 12
30 Oct 2007 - Main.KdDerr
META TOPICPARENT
name="VacuumEvaporator"
Glow Discharge in Edwards 306
Changed:
<
<
Recommended that GLow Discharge be carried out by the dedicated units
>
>
Recommended that Glow Discharge be carried out by the dedicated units
You will need to install the plasma ring for this procedure (see poster)
Load sample
Press
cycle
allow vacuum pump down to 1.0x 10
-1
Press
seal
If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
Turn voltage knob to HT
Turn Selector to 5
Allow plasma glow for 1min
Return selector and voltage knob both to 0
Press
seal
to exit
Vent
Visit Cryo EM Web site at
http://www.nysbc.org/facilities/CEM
--
DavidStokes
- 18 Mar 2005
Set ALLOWTOPICCHANGE =
CemfacGroup
META FILEATTACHMENT
attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT
attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""
View topic
|
H
istory
:
r14
<
r13
<
r12
<
r11
|
More topic actions...
Copyright
© by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet,
Send feedback