Glow Discharge in Edwards 306


  • Recommended that Glow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
Changed:
<
<
glowdischarge.jpg
>
>
glowdischarge.jpg
 

  • Load sample
  • Press cycle allow vacuum pump down to 1.0x 10-1
  • Press seal
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 90 Seconds
  • Return selector and voltage knob both to 0
  • Press seal to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

Deleted:
<
<
-- DavidStokes - 18 Mar 2005
 

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""

Revision 1316 Jun 2008 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that Glow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster) glowdischarge.jpg

  • Load sample
  • Press cycle allow vacuum pump down to 1.0x 10-1
  • Press seal
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
Changed:
<
<
  • Allow plasma glow for 1min
>
>
  • Allow plasma glow for 90 Seconds
 
  • Return selector and voltage knob both to 0
  • Press seal to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""

Revision 1230 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


Changed:
<
<
  • Recommended that GLow Discharge be carried out by the dedicated units
>
>
  • Recommended that Glow Discharge be carried out by the dedicated units
 
  • You will need to install the plasma ring for this procedure (see poster) glowdischarge.jpg

  • Load sample
  • Press cycle allow vacuum pump down to 1.0x 10-1
  • Press seal
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press seal to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""

Revision 1130 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster) glowdischarge.jpg

  • Load sample
Changed:
<
<
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
>
>
  • Press cycle allow vacuum pump down to 1.0x 10-1
  • Press seal
 
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
Changed:
<
<
  • Press "seal" to exit
>
>
  • Press seal to exit
 
  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""

Revision 1030 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
Changed:
<
<
* glowdischarge.jpg:
>
>
  glowdischarge.jpg

  • Load sample
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""

Revision 930 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
Changed:
<
<
>
>
* glowdischarge.jpg:
Added:
>
>
glowdischarge.jpg
 
Added:
>
>
 
  • Load sample
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Deleted:
<
<
* glowdischarge.jpg:
glowdischarge.jpg
 
Changed:
<
<
META FILEATTACHMENT attachment="BOCPOSTER.ppt" attr="" comment="Glow Discharge Conversion" date="1186412769" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" stream="BOCPOSTER.ppt" user="Main.KdDerr" version="0"
META FILEATTACHMENT attachment="glowdischarge.jpg" attr="" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" stream="glowdischarge.jpg" user="Main.KdDerr" version="1"
>
>
META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""
 

Revision 830 Oct 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
  • BOCPOSTER.ppt: Glow Discharge Conversion

  • Load sample
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>
* glowdischarge.jpg:
glowdischarge.jpg
 
META FILEATTACHMENT attachment="BOCPOSTER.ppt" attr="" comment="Glow Discharge Conversion" date="1186412769" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" stream="BOCPOSTER.ppt" user="Main.KdDerr" version="0"
Added:
>
>
META FILEATTACHMENT attachment="glowdischarge.jpg" attr="" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" stream="glowdischarge.jpg" user="Main.KdDerr" version="1"
 

Revision 706 Aug 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
Added:
>
>
 
  • Load sample
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Changed:
<
<
* BOCPOSTER.ppt: Glow Discharge Conversion
>
>
 
META FILEATTACHMENT attachment="BOCPOSTER.ppt" attr="" comment="Glow Discharge Conversion" date="1186412769" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" stream="BOCPOSTER.ppt" user="Main.KdDerr" version="0"

Revision 606 Aug 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)

  • Load sample
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>
* BOCPOSTER.ppt: Glow Discharge Conversion

META FILEATTACHMENT attachment="BOCPOSTER.ppt" attr="" comment="Glow Discharge Conversion" date="1186412769" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" stream="BOCPOSTER.ppt" user="Main.KdDerr" version="0"
 

Revision 506 Aug 2007 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
Changed:
<
<
  • You will need to install the plasma ring for this procedure
>
>
  • You will need to install the plasma ring for this procedure (see poster)
 
  • Load sample
Changed:
<
<
  • Pump down to 1.0x 10-1
>
>
  • Press "cycle" allow vacuum pump down to 1.0x 10-1
 
  • Press "seal"
Deleted:
<
<
  • Press "process" Allow display to go through rough and fine pumping sequence to "Plasma Process." This is a four minute sequence allowing the bleed valve to be opened.
 
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
Changed:
<
<
  • Press "seal" to exit "Plasma Process"
>
>
  • Press "seal" to exit
 
  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 409 Jun 2006 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Changed:
<
<
Glow Discharge in Edwards 306


  • You will not need any special accessories for this procedure

  • Load sample
  • Pump down to 1.0x 10-1
  • Press "seal"
  • Press "process" Allow display to go through rough and fine pumping sequence to "Plasma Process." This is a four minute sequence allowing the bleed valve to be opened.
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit "Plasma Process"

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

>
>
Glow Discharge in Edwards 306


  • Recommended that GLow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure

  • Load sample
  • Pump down to 1.0x 10-1
  • Press "seal"
  • Press "process" Allow display to go through rough and fine pumping sequence to "Plasma Process." This is a four minute sequence allowing the bleed valve to be opened.
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
  • Press "seal" to exit "Plasma Process"

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Deleted:
<
<
 

Revision 322 Apr 2005 - Main.KdDerr

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • You will not need any special accessories for this procedure

  • Load sample
  • Pump down to 1.0x 10-1
  • Press "seal"
Changed:
<
<
  • If the vacuum is too good adjust with unmarked knob under bell jar
>
>
  • Press "process" Allow display to go through rough and fine pumping sequence to
Added:
>
>
"Plasma Process." This is a four minute sequence allowing the bleed valve to be opened.
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
 
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0
Changed:
<
<
>
>
  • Press "seal" to exit "Plasma Process"
Added:
>
>
 
  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Revision 222 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • You will not need any special accessories for this procedure

  • Load sample
  • Pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

Added:
>
>
 

Revision 118 Mar 2005 - Main.DavidStokes

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • You will not need any special accessories for this procedure

  • Load sample
  • Pump down to 1.0x 10-1
  • Press "seal"
  • If the vacuum is too good adjust with unmarked knob under bell jar
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 1min
  • Return selector and voltage knob both to 0

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

-- DavidStokes - 18 Mar 2005

 
Copyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding this intranet, Send feedback