Glow Discharge in Edwards 306
Recommended that Glow Discharge be carried out by the dedicated units
You will need to install the plasma ring for this procedure (see poster)
Load sample
Press
cycle
allow vacuum pump down to 1.0x 10
-1
Press
seal
If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
Turn voltage knob to HT
Turn Selector to 5
Changed:
<
<
Allow plasma glow for 1min
>
>
Allow plasma glow for 90 Seconds
Return selector and voltage knob both to 0
Press
seal
to exit
Vent