Difference: GlowDischarge (13 vs. 14)

Revision 1416 May 2012 - Main.EdEng

 
META TOPICPARENT name="VacuumEvaporator"
Glow Discharge in Edwards 306


  • Recommended that Glow Discharge be carried out by the dedicated units
  • You will need to install the plasma ring for this procedure (see poster)
Changed:
<
<
glowdischarge.jpg
>
>
glowdischarge.jpg
 

  • Load sample
  • Press cycle allow vacuum pump down to 1.0x 10-1
  • Press seal
  • If the vacuum is too good adjust with unmarked knob under bell jar (bleed valve)
  • Turn voltage knob to HT
  • Turn Selector to 5
  • Allow plasma glow for 90 Seconds
  • Return selector and voltage knob both to 0
  • Press seal to exit

  • Vent


Visit Cryo EM Web site at http://www.nysbc.org/facilities/CEM

Deleted:
<
<
-- DavidStokes - 18 Mar 2005
 

META FILEATTACHMENT attr="" autoattached="1" comment="" date="1193769600" name="glowdischarge.jpg" path="glowdischarge.jpg" size="267954" user="Main.KdDerr" version="1"
META FILEATTACHMENT attr="" autoattached="1" comment="Glow Discharge Conversion" date="1186412770" name="BOCPOSTER.ppt" path="BOCPOSTER.ppt" size="4520960" user="Main.KdDerr" version=""
 